![AND UM-211 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/and/um-211/um-211_instruction-manual_2941387016.webp)
English
15
The measurement
is started.
Auscultation
exhaust speed
changing mode
START
STOP
The measurement
is started.
Measurement
standby mode
Measurement
standby mode
Auscultation setting
1. Press the MODE button at the
pressurization setting mode to go into
auscultation setting mode. The
“
F02
”
is
displayed at the display for systolic
pressure, and the current status is
displayed at the display for diastolic
pressure
2. Press to
or
button to switch
between ON or OFF. The device
illuminates the auscultation mark when
the auscultation mode is set to ON.
3. Press the MODE button when the
auscultation mode is set to ON to
proceed to auscultation exhaust speed
changing mode.
Press the MODE button when the
auscultation mode is set to OFF to
proceed to measurement standby mode.
Press the START/STOP button to start
the measurement. Also, the device
proceeds to standby mode
automatically after a regular time.
Auscultation exhaust speed changing
Note: Select “Lo” when measuring normally. Should the patient pulse appear to
be 100 or higher, measuring at the “Hi” is possible.
1. Press the MODE button at the auscultation
setting mode when the auscultation setting
is set to ON to go into auscultation exhaust
speed changing mode.
The “F03” is displayed at the display for
systolic pressure, and the current status
is displayed at the display for diastolic
pressure
2. Press
to
or
button to switch
between Hi or Lo.
3. Press the MODE button to proceed to
measurement standby mode.
Press the START/STOP to start the
measurement. Also, the device
proceeds to standby mode
automatically after a regular time.
START
STOP