MONITOR COMPONENTS
E-2
?
LifeSource Health Line (Toll-Free): 1-888-726-9966
WHAT DISPLAY SYMBOLS MEAN
?
PULSE DISPLAY ERROR
❤
Display
Symbol
Recommended
Action
Remain still and refrain
from talking during
monitoring process.
Take measurement again
and consult with your
physician.
Replace batteries with
two new batteries.
Check and re-fasten cuff
for correct fit and take
measurement again.
Take another reading,
making sure you remain
quiet and perfectly still.
Check cuff to ensure fit
and fastening is correct,
and measure again.
Check for correct fit and
position of cuff and take
another reading.
Condition/
Error
Appears when measuring
is in progress, and flashes
when pulse is detected.
An irregular heartbeat or
body movement may
have occurred. Refer to
page E-7 for more infor-
mation on irregular
heartbeats.
Appears when battery
voltage is insufficient for
device to function.
Appears when systolic
and diastolic measure-
ments are within 10
mmHg of each other.
Appears when the pres-
sure reading is unstable
due to movement during
the measurement.
Appears when the cuff is
not fastened correctly.
Appears when pulse is
not sufficiently detected.
Indicates previous
measurement stored in
memory.
LOW BATTERY
?
(Example data display)
Case
Cuff
Battery holder
cover
Display
Systolic pressure
Diastolic pressure
Pulse
MEMORY button
START button
UB328_0104 2/3/04 10:51 AM Page 6