Dismounting and reassembling
AS 21hp, AS 10e
16
3.2
Exchange of Z-drive
Move the sampler into the service position.
Fig. 17 Unplugging the electrical connector from the main board (exchange of Z-drive)
Unplug the Z-sensor cable (X10) and the Z-motor cable (X6).
Loosen the 2 screws and remove cable holders.
Fig. 18 Loosening the 3 screws of the Z-drive