To use your dry contact output fail relay you must connect a 9-pin D connector to the unit.  

You must use pins one and six as is indicated on page 7 in the remote connector diagram.

The relay is factory preset to fail in the closed position when the low output LED and buzzer 

come on.  If you wish to have the relay fail in the open position when the low output LED 

and buzzer come on, you must take the cover off the unit and move the jumper on J10 to the 

other position.  J10 is located next to the relay K1.

To change the position of the jumper, first turn the unit off and disconnect the unit from both 

the power and load(s).  Next, turn the unit on for 30 seconds to discharge the capacitors, 

then turn it off again.  Remove the eight screws holding on the cover.  Turn the unit upside 

down, remove the cover and locate J10.  It will be next to the relay K1 as is shown in the 

above diagram.  Simply move the jumper to the desired position as is shown in the above 

diagram.  Replace the cover and re-install the eight cover screws.  Reconnect the unit to the 

power and load(s).

Dry Contact Relay

Meters

High quality digital meters can be added to the charger (factory installed only).  Both outputs 

can be monitored for current and voltage.

The units may be configured for load sharing if they are equiped with the optional output 

isolation diodes. To confirm that your unit has these diodes, use an ohmmeter to measure the 

resistance between the two positive output terminals. If the diodes are present the terminals 

will measure as not connected. If the diodes are NOT present, the terminals will measure as 

a short circuit. Assuming the output isolation diodes are present, connect a one-foot piece of 

red wire of the appropriate gauge (as shown in the table below)

LOAD SHARE OPTION

OUTPUT 1

OUTPUT 1

OUTPUT 2

OUTPUT 2

Содержание PWS1000

Страница 1: ...00 Power Supply 8128 River Way Delta B C V4G 1K5 Canada T 604 946 9981 F 604 946 9983 TF 800 668 3884 US CANADA www analyticsystems com An ISO9001 and AS9100 Registered Company Battery Chargers Inverters Power Supplies Voltage Converters ...

Страница 2: ...Copyright 2005 2013 Analytic Systems Ware 1993 Ltd ...

Страница 3: ...ND AC POWER CORD CONNECTION INSTRUCTIONS The plug must be plugged into an outlet that is properly installed and grounded in accordance with al local codes and ordinances DANGER Never alter AC cord or plug provided if it will not fit outlet have proper cord installed by a qualified electrician Improper connection can result in a risk of an electric shock Analytic Systems does not recommend the use ...

Страница 4: ...or 220 VAC power source High quality digital meters can be added factory option to allow monitoring of charging current and charging voltage Other output voltages up to 72 volts are also available by special order Specifications Input Voltages Nominal ip 110Vac 220Vac Actual 90 130Vac 180 260Vac Input Amps max 13 2 12V 17 24 32 48V 6 6 12V 8 5 24 32 48V Input Fuse slow blow MDA 20 12V MDA 25 24 32...

Страница 5: ... V4G 1K5 Canada p 604 946 9981 f 604 946 9983 tf 800 668 3884 US Canada www analyticsystems com analyticinfo analyticsystems com Revised January 2014 Specifications subjects to change without notice Installation MOUNTING Mount the unit in a DRY location Allow at least 4 inches of clearance around the heat sink fins for adequate cooling POWER CONNECTION The unit is supplied with a 5 foot power cabl...

Страница 6: ...t When you are done replace the cover plate and securely tighten the screws All connections should be made inside an appropriate junction box The maximum current draw from the 110 VAC supply is 14 6 amps so a 20 amp circuit breaker should be used in the circuit panel and for a 220 VAC supply 7 3 amps is the maximum current draw so a 10 amp circuit breaker should be used in the circuit panel to fee...

Страница 7: ...ver and locate J10 It will be next to the relay K1 as is shown in the above diagram Simply move the jumper to the desired position as is shown in the above diagram Replace the cover and re install the eight cover screws Reconnect the unit to the power and load s Dry Contact Relay Meters High quality digital meters can be added to the charger factory installed only Both outputs can be monitored for...

Страница 8: ...cating all the diagnostic indicators and audible alarm REMOTE CONNECTOR This connector is located on the side of the unit Important To prevent the possibility of High Voltage Electrical Shock do not power up the Power Supply unless all wiring from the unit to the remote is securely connected Do not remove the dust cover from the DB 9 connector if the remote is not being used Remote Control Option ...

Страница 9: ...e LOW INPUT Indicates that the input voltage is below normal because The input voltage is not in the correct range for proper operation of the unit OVERTEMP Indicates that the Battery Charger is running too hot because Too much power is being drawn turn off or unplug some devices The Battery Charger is located in a poorly ventilated area OVERLOAD Indicates that the load it trying to draw to much c...

Страница 10: ...10 ...

Страница 11: ...ing will incur additional charges and are the responsibility of the Purchaser 6 Analytic Systems reserves the right to void the warranty if a Labels identification marks or serial numbers are removed or altered in any way b Our invoice is unpaid c The defect is the result of misuse neglect improper installation environmental conditions non autho rized repair alteration or accident 7 No refund of t...

Страница 12: ... Way Delta B C V4G 1K5 Canada T 604 946 9981 F 604 946 9983 TF 800 668 3884 US CANADA www analyticsystems com An ISO9001 and AS9100 Registered Company Battery Chargers Inverters Power Supplies Voltage Converters ...

Отзывы: