Sub MkIIIF
User Manual
Int. Approved
Document ref: ASF3-800-08
November 2018
Page 78 of 79
Copyright © 2009 Analox Ltd. All Rights Reserved.
Commercial in Confidence
CO
2
Sensor
Analox low power, long life infra-red sensor.
CO2 reading is pressure compensated by microprocessor
Ranges
0 to 20mBar, 0 to 100mBar (enquire regarding alternative
ranges)
Accuracy
at atmospheric pressure and within 0-50%FS :
±2%FS,
±0.1%FS/°C
at atm pressure and within 50-100%FS : ±3%FS, ±0.1%FS/°C
0.8< pressure < 6.0 BarA, CO2 < 50% FS : ±3%FS, ±0.1%FS/°C
otherwise : ±5%FS, ±0.1%FS/°C
Choice of units from mBar, %, %SEV, ppm, ppmSEV and mATA (define at
time of order, please enquire regarding alternative units)
Temperature
Sensor
Analox OEM HAT sensor
Range
-10
o
C to +60
o
C
Accuracy
±1°C ± 1display digit
Choice of units in either
o
C or
o
F (define at time of order)
Humidity Sensor
Analox OEM HAT sensor
Range
0-100% RH
Accuracy
±10%RH
Analogue
Outputs
8x 4-20mA output channels
–
(active outputs
–
generating current output)
Voltage outputs (0-10V) also available by request.
Relay Outputs
8x Relay outputs assignable to any alarm or fault conditions. Changeover
contacts on each relay, normally configured in Fail-Safe Mode.
Contacts rated at load 1 Amp (24 VDC),
Minimum load 1mA at 5V DC
Operating
Temperature
-10
o
C to +40
o
C
Storage
Temperature
-40
o
C to 66
o
C (remove oxygen sensor(s) for storage below -15
o
C)
Dimensions
Operator Console
240mm (w) x 133mm (h) x 80mm (d)
Remote Sensor REM1/2/3
150 x 150 x 85mm (whd)
Weight
Operator Console
2.5kg
REM1
1.8kg
REM2
1.6kg
REM3 1.6kg
Materials
Operator Console : SS
Remote Sensors : SS
Содержание Sub Mk III F