EII CO Technical User Manual 

 

Document Ref: P0107-801-00 - February 2011 
Page 6 

 

Introduction 

1.1 

Scope of this Manual 

This Technical manual is supplementary to the EII CO User Manual (P107-800). This manual 
details  the  procedure  for  performing  a  zero  and  span  calibration  of  the  EII  CO.  For  general 
details of using and maintaining the EII CO please refer to the EII CO user manual 
 
 

 

 

1.2 

About the EII CO 

The  EII  CO  carbon  monoxide  analyser  is  designed  to  measure  CO  levels  in  the  range  0  to 
50ppm for tank CO level checking.  
 
The EII CO is ergonomically designed, and equipped with several features to ensure ease of 
use and reliability. The instrument has been  designed to be  held in the left hand to facilitate 
checking  your  tank.  It  is  fitted  with  a  large  digital  display  and  incorporates  an  internal 
electrochemical  carbon  monoxide  sensor.    Power  is  provided  by  a  9V  battery  which  will  last 
for  approximately  1  year  before  replacement  is  necessary.  The  EII  CO  will  automatically 
switch  off  after  10  minutes  to  ensure  battery  life  is  not  compromised  if  the  instrument  is 
accidentally turned on.  
 
The EII CO is water and drop resistant. Designed specifically for the diving industry – whether 
you may be a Sport, Commercial or Military diver - where hostile environmental conditions are 
the norm not the exception.  
 

Содержание EII CO

Страница 1: ...EII CO Technical User Manual Analox Ltd 15 Ellerbeck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net...

Страница 2: ......

Страница 3: ...Ref P0107 801 00 February 2011 Page 4 List of Contents 1 Introduction 6 1 1 Scope of this Manual 6 1 2 About the EII CO 6 2 Calibration 7 2 1 Items required 7 2 2 Calibration procedure 7 3 Safety Inf...

Страница 4: ...EII CO Technical User Manual Document Ref P0107 801 00 February 2011 Page 5...

Страница 5: ...esigned and equipped with several features to ensure ease of use and reliability The instrument has been designed to be held in the left hand to facilitate checking your tank It is fitted with a large...

Страница 6: ...a flat blade screwdriver loosen the 4 M2 5 slotted screws located on the front cover these are captive screws so can not be fully removed from the cover The gasket seal is designed to prevent water le...

Страница 7: ...e screws and black fibre washers Remove the rear cover 5 Screw the flow indicator on to the zero calibration cylinder synthetic air and fit the flow adaptor to the tubing Flow indicator fitted Flow ad...

Страница 8: ...to read 000 Display set to 000 9 Close the control valve of the flow indicator Remove the control valve from the zero calibration cylinder and fit it to the span calibration cylinder 10ppm carbon mono...

Страница 9: ...Place the lanyard loop over the dowel pin then fit the back cover in place making sure the CO sensor lead runs down the rear of the battery compartment as shown Affix the cover using the two M3x16mm...

Страница 10: ...2011 Page 11 15 Feed the CO sensor cable so that it curls around the inside of the middle section fit the red front section Secure in place with the four M2 5 slotted screws taking care not to trap or...

Страница 11: ...se Obtain medical advice if continued irritation Ingestion Corrosive May cause sore throat abdominal pain nausea and severe burns of the mouth throat and stomach and may be fatal If swallowed DO NOT I...

Страница 12: ...Bump test gas kit 10ppm CO in balance air Providing around 20 bump tests from a standard bump test kit SA7CAN510 Span calibration gas cylinder 10ppm carbon monoxide in balance air SA7EIICOCALKIT Zero...

Страница 13: ...P0107 801 00 February 2011 Page 14 5 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for information on the disp...

Отзывы: