HARDWARE CONFIGURATION 

To configure the hardware, follow these steps: 

1.

 

Plug the USB A to Mini-B cable into the PC, if the cable is 
not already connected.  

2.

 

Connect the ISEB to the USB cable.  
If prompted to install the device, see the Software 
Configuration section for details on how to install and 
select the correct driver. 

3.

 

Connect the ISEB to the ADXL345 satellite using the  
18-inch, 20-pin ribbon cable. This cable is keyed to prevent 
inserting it backwards and causing damage to the system.  

4.

 

Connect the ADXL345 satellite to the opposite end of the 
ribbon cable.  

5.

 

Verify that the jumpers located in the center of the 
ADXL345 satellite are configured as shown in Figure 5. 

CO

NN

E

CT

O

R

ADXL345

JUMPER

POSITIONS

08

11

9-

0

05

 

Figure 5. Correct Jumper Positions for the ADXL345 Satellite  

6.

 

Place the ADXL345 ultralow power accelerometer into the 
socket on the satellite board.  
The Socket Pin 1 Indicator is located inside the socket close 
to the hinge, as shown in Figure 6. This pin indicator should 
match up with the Pin 1 Indicator on the ADXL345.  

7.

 

After positioning the ADXL345 in the socket, firmly close 
the socket until it latches. 

ADXL345 SOCKET
PIN 1 INDICATOR

ADXL345
ACCELEROMETER
PIN 1 INDICATOR

08

11

9-

00

6

 

Figure 6. Location of Pin 1 Indicators for the ADXL345 Accelerometer and Socket 

The inertial sensor evaluation system should now be set up and 
ready to use. 

 

Содержание iMEMS ADXL345

Страница 1: ...s directly to a PC via its USB where the USB connection provides both power and communications to the board The ISEB then connects to the satellite board through a ribbon cable This cable allows the satellite to be easily manipulated for testing or for being separately placed into an environmental chamber for temperature or humidity testing Separating the boards mitigates corruption of data due to...

Страница 2: ...Setup 3 Software Configuration 3 Hardware Configuration 4 Updating Repairing the ISEB Firmware 5 Inertial Sensor Evaluation System Graphical User Interface 6 Configuration Tab 6 Real Time Measurement Tab 7 Interrupt Configuration Tab 8 Power Consumption Tab 9 Temperature Tab 10 Header Pinout 11 ESD Caution 11 REVISION HISTORY 6 09 Revision 0 Initial Revision ...

Страница 3: ...see Figure 2 08119 002 Figure 2 Found New Hardware Prompt 3 Select Don t search I will choose the driver to install see Figure 3 and select ADI Inertial Sensor Evaluation System from the model list see Figure 4 The ISEB should be detected automatically in the Device Manager as the ADI Inertial Sensor Evaluation System under the Ports COM LPT selection It is recommended to open the Device Manager t...

Страница 4: ...e ribbon cable 5 Verify that the jumpers located in the center of the ADXL345 satellite are configured as shown in Figure 5 CONNECTOR ADXL345 JUMPER POSITIONS 08119 005 Figure 5 Correct Jumper Positions for the ADXL345 Satellite 6 Place the ADXL345 ultralow power accelerometer into the socket on the satellite board The Socket Pin 1 Indicator is located inside the socket close to the hinge as shown...

Страница 5: ...hen the COM port is selected click OK to accept the changes and go back to the main screen 08119 108 Figure 8 Selecting the Correct COM Port for the Downloader When the ISEB is connected the correct firmware is selected and the downloader is fully configured follow these steps to flash the firmware 1 Click Start as shown in Figure 7 to initiate the flashing process After clicking Start the two but...

Страница 6: ...I Inertial Sensor Evaluation System under the Ports COM LPT selection 08119 007 Figure 10 Selecting the Correct COM Port for the ISEB After selecting the COM port click on Connect to connect to the ISEB This causes the indicator under Vs Set to light up and update the Current Vs V value verifying that the board is connected see Figure 11 08119 008 Figure 11 ISEB Connected to PC Clicking the round ...

Страница 7: ...licking the Self Test button when self test is enabled turns the self test feature off and causes the Self Test indicator to dim Self test is enabled as long as the indicator is lit The ADXL345 provides three offset registers that are used for calibration and offset adjustment On the Real Time Measurement tab these registers are easily accessible for programming offset values This can be done in t...

Страница 8: ...ils refer to the ADXL345 data sheet Clicking an indicator causes it to toggle values A lit indicator causes a value of 1 to be written to that bit and an unlit indicator causes a value of 0 to be written to that bit In the Interrupt Configuration box you can select the interrupts to enable For purposes of the GUI all enabled interrupts are mapped to the INT2 pin Selecting the checkbox next to the ...

Страница 9: ...n indicator lights up signifying that a controller or microprocessor is actively pulling data from the part or processing it In addition to power savings from the ADXL345 more power can be saved by powering down the microprocessor during periods of inactivity and using the ADXL345 activity interrupt to wake the microprocessor up during periods of activity The ACT INACT_CTL indicators can be used t...

Страница 10: ... Figure 15 allows a maximum number of points to be set and at what rate data is acquired The testing can be stopped before reaching the maximum number of points The data rate for the accelerometer can also be set to ensure that the testing is performed under conditions most like those that the part would be configured for in a system Figure 15 Temperature Tab The data in the top graph in Figure 15...

Страница 11: ... HEADER PINOUT 1 S1 2 3 S2 4 5 S3 6 7 S4 8 CS2 9 VS2 10 VS1 11 12 ST1 13 ST2 14 SCL1 15 SDA1 16 CLK 17 MOSI 18 MISO 19 CS3 20 CS4 P1 3 3V CS2 VS2 VS1 ST1 ST2 SCL1 SDA1 CLK MOSI MISO CS3 CS4 AGND GND 08119 013 Figure 16 ISEB 20 Pin Header Pinout ESD CAUTION ...

Страница 12: ... use of evaluation boards Information furnished by Analog Devices is believed to be accurate and reliable However no responsibility is assumed by Analog Devices for its use nor for any infringements of patents or other rights of third parties that may result from its use Analog Devices reserves the right to change devices or specifications at any time without notice Trademarks and registered trade...

Отзывы: