1-1
Safety Precautions
Operator Manual
150830-026
The following
Safety Precautions
must
be observed when operating and servicing this equipment. WARNING indicates
the potential for personal injury, and CAUTION indicates the potential for damage to equipment. These
Safety
Precautions
are repeated, where applicable, throughout the manual.
WARNING – PINCHING HAZARD:
Pinch points are created during extreme tabletop articulation. Carefully review illustrations in Figure 2-1 before
operating the table.
WARNING – TIPPING HAZARD:
Do not place patient on the table unless floor locks are engaged.
Do not release floor locks while patient is on table.
Do not use this table for patients exceeding the 1,000-lb (452-kg) limit when patient is positioned in normal
orientation. The maximum safe patient weight on this table for standard surgical positions in normal orientation is
1,000 lb (452 kg) with floor locks locked.
Do not use this table for patients exceeding the 500-lb (226-kg) limit when patient is positioned in reversed
orientation. The maximum safe patient weight on this table for standard surgical positions in reversed orientation
is 500 lb (226 kg) with floor locks locked.
When performing surgery requiring a headrest accessory in reversed patient orientation, or when using a Fem/Pop
board or the 3080/3085 Ortho Extension accessory, do not exceed the 400 lb (181 kg) patient weight.
Do not use the Fem/Pop Board with X-ray Tops for bariatric patients.
Foot Extension Accessory or combination of Foot Extension and Headrest Accessories from previous design
STERIS tables must not be used for reverse orientation on the 3085 SP Table.
Do not use two or more Uro-Endo/Image Amplification Extension accessories together on the 3085 SP Table.
Do not articulate table with auxiliary override systems unless floor locks are engaged.
During an articulation if the tabletop sections contact an obstruction, the table may tip. Before lowering either the
table top or individual sections, remove possible obstructions. Do not allow leg section, when lowered, to contact
the floor.
Fem/Pop Board must be installed into leg section only. Board must be used to support the legs only. It is not
intended to support upper body weight.
Do not use the Fem/Pop Board with patients exceeding 400 lbs (181 kg).
WARNING – EXPLOSION HAZARD:
Table must not be used in the presence of flammable anesthetics.
WARNING – TRIPPING HAZARD:
Route the power cord to the receptacle in a position so it will not be tripped over by personnel in the area.
SAFETY PRECAUTIONS
1