The last four addresses of the FW Data 2 section contain the checksums which are stored to 
GP30 when downloading firmware to GP30. These fields are directly updated, when firmware 
files are loaded or content of firmware data fields are changed. 

 

Pressing 

the “Verify FW” button after downloading compares the content of the NVRAMs with 

the given files by their checksums. The software calculates the checksum of the given files and 

reads the calculated checksums of GP30 as well as the stored checksums at the end of FWD2 
section. Note: The firmware data file word 127 is by default empty, not knowing the checksum of 
the on-chip 

ams

 firmware. 

Figure 26: 
Firmware ams Code 

 

 

So copy manually the calculated checksum for the 

ams

 

code into the field “Checksum FWY manual 

entry”. Word 127 in the data will be updated and after downloading again the verification will pass for 
all. 

Figure 27: 
Verification by Calculated Checksum 

 

 

 

In the “Firmware ams Code” section, the checksums for the 

ams

 firmware code are also 

checked and displayed after a “Verify FW”. The 

ams

 firmware code cannot be modified by user. 

Therefore a checksum calculated by software field is missing in this section. 

 

A lock state of GP30 or a hang-up, caused by a faulty firmware user code can be dissolved by 
pressing “Erase FW” button. After that, a new firmware (user code & data) need to be 
downloaded again. 

 

CPU Values 

This tab is only for customer who uses the ams firmware for flow calculation.

 It reads out some 

important CPU values like water temperature, flow, velocity, etc. 

To enable the readout the “Read 

calculated values” checkbox has to be set. 

The lower sections allows to read from any RAM addresses. 

 

 

Содержание GP30-DEV-KIT

Страница 1: ...User Guide UG000292 TDC GP30 Reference Board GP30 DEV KIT v1 00 2019 Feb 26...

Страница 2: ...duction 8 3 2 Communication Interface 8 4 Software Description 10 4 1 Measurement 10 4 2 Ultrasonic Measurement Control 13 4 3 Temperature Measurement Control 14 4 4 General Control 15 4 5 Interfaces...

Страница 3: ...pment kit offers user friendly configuration and extensive testing of the TDC GP30 but also the complete assembler environment for programming the device For a proper use of the evaluation system we s...

Страница 4: ...Description GP30 DEV KIT 220260003 TDC GP30 Demo Kit for QFN32 version including PICOPROG and cables GP30 DEMO KIT F01 220260006 TDC GP30 Demo Kit with firmware for QFN32 version including PICOPROG a...

Страница 5: ...he package to the desired directory Open setup exe from the unzipped directory Follow the instructions on the screen 2 2 Install the Hardware Make sure software is installed correctly before proceedin...

Страница 6: ...d Comm With GP30 OK should become green Next open our configuration GP30Y_config_default_A1 A2 11 03 and download it into the chip pressing Write Config Connect your spool piece to pins US_UP and US_D...

Страница 7: ...Program Files x86 ams GP30 data and install the driver for your operating system manually In case of an upgrade of the software to a newer version please make sure that the software uses the latest dr...

Страница 8: ...ports of TDC GP30 are available Additional patch fields allow an easy extension with additional circuits Those can be amplifiers or analog switches for operation in gas meter mode Figure 5 GP30 DEMO M...

Страница 9: ...ICOPROG is now listed as UNIPRO in the device manager For SPI communication only PICOPROG version 2 0 is sufficient Figure 6 shows the connection between the PICOPROG and the GP30 board The flat conne...

Страница 10: ...ts tab Figure 7 Main Window 4 1 1 First Step with Measurement Control Elements A good first step is to load a working configuration and make measurements in frontend mode without using the internal 32...

Страница 11: ...ely and also the sum of all measured hits These 9 results are displayed for both directions as we call them up and down The evaluation software additionally calculates the difference between up and do...

Страница 12: ...activate up to four plots Each plot has various selections e g TOF1UP TOF2UP etc Always averaged values are displayed Figure 9 TOF Graph The measurement data can be exported into text files either th...

Страница 13: ...t results opens in a separate window by Temperature Graph button similar to TOF graph Figure 12 Temperature Measurement Results Graph 4 2 Ultrasonic Measurement Control In this tab the user makes all...

Страница 14: ...meanings of the various settings are displayed in clear text For more details about the register settings please refer to the GP30 manual 4 3 Temperature Measurement Control All settings for an appro...

Страница 15: ...surement Control Tab The meanings of the various settings are displayed in clear text For more details about the register settings please refer to the GP30 manual 4 4 General Control The General Contr...

Страница 16: ...ails about the register settings please refer to the GP30 manual 4 4 1 Firmware In case the TDC GP30 has firmware setting flag Enable post processing turns on the CPU flow meter mode Having this not s...

Страница 17: ...se refer to the GP30 manual 4 5 1 Pulse Interface The pulse interface needs an appropriate firmware in the chip The settings in the evaluation software only generate an artificial pulse to test the ge...

Страница 18: ...s about the register settings please refer to the GP30 manual 4 7 ams On this tab named acam specific parameters are displayed but for ams internal use and analysis only 4 8 Firmware Parameters and Fi...

Страница 19: ...Document Feedback TDC GP30 Software Description Demo Kit Manual PUBLIC UG000292 v1 00 2019 Feb 26 40 19 Figure 18 Firmware Parameters Tab Figure 19 Firmware Error Counters...

Страница 20: ...cessary for a proper configuration of the GP30 After opening this file the control settings are updated in the GUI Save Config This menu item allows the saving of the current GUI control settings into...

Страница 21: ...Firmware Download window Opens a window which shows the registers important for a proper configuration setting of the GP30 In the left column the register contents correspond to the settings done in...

Страница 22: ...ion CR_ and the system handling SHR_ register CPU Init Clears the CPU block in GP30 SV Init Clears the supervisor block in GP30 FEP Init Clears the frontend processing block in GP30 Request Release Bu...

Страница 23: ...Document Feedback TDC GP30 Software Menu Demo Kit Manual PUBLIC UG000292 v1 00 2019 Feb 26 40 23 Figure 23 Firmware Menu The following window comes up...

Страница 24: ...options The download option effects that Firmware Download window is opened see also below Whether the call of these functions was successful or not is indicated by the messages at the bottom of the...

Страница 25: ...tool and intended for the user part of 4kx8 Program NVRAM can be loaded by pressing Open File In the Firmware Data section a firmware data file dat which is intended for the 128x32 Data NVRAM can be...

Страница 26: ...anually the calculated checksum for the ams code into the field Checksum FWY manual entry Word 127 in the data will be updated and after downloading again the verification will pass for all Figure 27...

Страница 27: ...ure 28 Flow Calculation Tab 5 4 Help When moving the cursor over the values in tabs of main window the parameter name used in the GP30 manual is displayed By right click and selection of Description a...

Страница 28: ...26 40 28 Figure 29 Description and Tip Window Help Contents Not supported in this software revision USB Communication Figure 30 USB Communications As described in chapter Software Installation About D...

Страница 29: ...Document Feedback TDC GP30 Schematics Layers and BOM Demo Kit Manual PUBLIC UG000292 v1 00 2019 Feb 26 40 29 6 Schematics Layers and BOM Figure 31 GP30 DEMO MODULE Schematics...

Страница 30: ...XC6206 Voltage regulator 3 1 X1 4 MHz CSTR_G Ceramic resonator Murata 4 1 X2 32 768 kHz KX 327XS Quartz crystal Geyer 5 1 C1 4 7 F C805 Chip capacitor 6 2 C2 C3 100 F F95_P Solid tantalum 7 1 C4 100 n...

Страница 31: ...R8 10 M R805 Chip resistor 16 1 R9 1 k R805 Chip resistor 17 1 R10 560 k R805 Chip resistor 18 1 J13 2 pol ST 254_2 Connector for power supply combined with J22 19 1 J22 7 pol ST 254_7_1R Connector f...

Страница 32: ...c and come un calibrated as the final housing will have a major impact on the calibration 7 1 1 Water Meter Modules Made of Brass DN15 and DN20 modules come with tubes made of brass and the sensor fro...

Страница 33: ...anent Flow Rate Q3 m h 2 500 4 000 6 300 10 000 16 000 Overload Flow Rate Q4 m h 3 125 5 000 7 875 12 500 20 000 Flow Range Ratio R Q3 Q1 R250 1 R400 Accuracy Class Class 2 Temperature Class T30 1 T50...

Страница 34: ...cteristics Nominal Diameter mm 15 20 25 32 40 Starting Flow m h 0 002 0 003 0 003 0 005 0 005 Minimum Flow Rate Q1 m h 0 010 0 016 0 025 0 040 0 064 Transitional Flow Rate Q2 m h 0 016 0 025 0 040 0 0...

Страница 35: ...UGM 2 YYYYMMDD SSSSSS 1 YYYY year MM month DD day SSSSSS product serial number that day Figure 42 Measurement Characteristics Application Residential Commercial Industrial Flow Range G1 6 to G4 G6 to...

Страница 36: ...le 86 177 5310 5044 Email wangchuan ieslab cn Address No 1677 Kehang Rd Licheng Dist Jinan City Shandong PR China WeChat wangchuan9418 7 2 Transducers For transducers we recommend products from Zhejia...

Страница 37: ...Vpp 110mm 20 6 0 2 10 2 0 2 16 8 0 2 7 6 0 2 PSC1 0M019168H2AD2 B0 f 1 0 0 1 MHz R1MHz 150 to 400 Cp 1000 pF 20 Top 40 to 85 C S 700mV 2Vpp 110mm PEEK PSC1 0M020160H2AD1 B0 f 1 0 0 1 MHz R1MHz 150 to...

Страница 38: ...150 to 500 Cp 1100 pF 20 Top 40 to 85 C S 200mV 3Vpp 140mm PSC4 0M018223H2AD2 B0 f 4 0 0 2 MHz R1MHz 30 to 150 Cp 2050 pF 20 Top 40 to 85 C S 200mV 3Vpp 140mm 1 Tstrg 40 to 85 C for all parts Contact...

Страница 39: ...ision Information Changes from previous version to current revision v1 00 Page Section 7 Reference Modules and Transducers added 31ff Page and figure numbers for the previous version may differ from p...

Страница 40: ...in connection with or arising out of the furnishing performance or use of the technical data herein No obligation or liability to recipient or any third party shall arise or flow out of ams AG renderi...

Отзывы: