Name
Range
Unit
Description
APS Presets
Min
Typ
Max
APS Presets
2
3
4
-
The number of presets assigned
to APS. Selected Presets will be
used for APS. Other presets are
presumed to be assigned for
other purposes (for example
monitor mode)
Threshold
0.1
0.25
6.0
dB
The threshold that must be
exceeded before switching
presets
HoldTime
0.5
1.0
10.0
seconds
The time, after entering a new
preset, when the algorithm can
consider changing preset.
Band
weighting 1
-60
0.0
0.0
dB
Manual weighting for frequency
band 1. Suppl
ements ‘A’
weighting and passive
weighting.
Band
weighting 2
-60
0.0
0.0
dB
Manual weighting for frequency
band 2. Supplements ‘A’
weighting and passive
weighting.
Band
weighting 3
-60
0.0
0.0
dB
Manual weighting for frequency
band 3. Supplements ‘A’
weighting and passive
weighting.
Band
weighting 4
-60
0.0
0.0
dB
Manual weighting for frequency
band 4. Supplements ‘A’
weighting and passive
weighting.