SYMBOLS
�
Caution ! Risk of electric shock
�
Caution ! Refer to the explanation in this Manual
�
Double Insulation or Reinforced insulation
�
Alternating Current (AC).
�
Direct Current (DC).
Fuse
�
Earth (Ground)
�
Please remove all the test leads before preforming maintenance,
cleaning, battery replacement, fuse replacement, etc
�
Complies with European Directives
�
Conforms to relevant Australian standards
=
Do not dispose of this product as unsorted municipal waste.
Contact a qualified recycler for disposal
Safety Information
This manual contains information and warnings that must be followed for
operating the instrument safely and maintaining the instrument in a safe
operating condition. If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the instrument may be impaired.
The meter is intended only for indoor use.
The meter protection rating, against the users, is double insulation per
IEC61010-1 2nd Ed., EN61010-1 2nd Ed., UL61010-1 2nd Ed. and CAN/CSA C22.2
No. 61010.1-0.92 to Category III 1000 Volts AC & DC and Category IV 600 Volts
AC & DC.
Measurement category:
V :
Category III 1000 Volts AC & DC, and Category IV 600 Volts AC & DC.
A / mAµA :
Category III and Category IV 600 Volts AC and 300 Volts DC.
Per IEC61010-1 2nd Ed. (2001) Measurement Category