14
Pollution degree:
2
Storage temperature:
-20°C to 60°C (-4°F to 140°F), < 80% R.H.
(With battery removed)
Altitude:
Operating below 2000m
Temperature coefficient:
Nominal 0.15 x (specified accuracy)/ °C @ (0°C -
18°C/32°F – 64.4°F) or 28°C - 45°C/82.4°F - 113°F),
or otherwise specified
Sensing:
AC, AC+DC true RMS
Safety:
Double insulation per iec61010-1 2nd ed.,
En61010-1 2nd ed., Ul61010-1 2nd ed. & Can/csa
c22.2 No. 61010.1-0.92 To category III 1000v AC
& DC and category IV 600V AC & DC
Overload protections:
µA & mA:
1A/600V, IR 10kA or better, F fuse
A:
10A/600V, IR 100kA or better, F fuse
V:
1050VRMS, 1450VPEAK
mV,
, & others:
600 VDC & VAC RMS
TERMINALS (TO COM) MEASUREMENT CATEGORY:
V:
CATEGORY III 1000V AC & DC AND CATEGORY
IV 600V AC & DC
A / mAµA:
CATEGORY III AND CATEGORY IV 600
VAC AND 300 VDC.
TRANSIENT PROTECTION:
8KV (1.2/50µS SURGE)
E.M.C.:
Meets en61326-1:2006 (EN55022, EN61000-3-
2, EN61000-3-3, EN61000- 4-2, EN61000-4-3,
EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4- 8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Power supply:
Single standard 9V battery NEDA1604, JIS006P
OR IEC6F22
Power consumption:
4.3mA typical
Low battery:
Below approx. 7V
Apo timing:
Idle for 17 minutes
Apo consumption:
50µA typical