c
RESISTANCE / CONTINUITYMEASUREMENT
CAUTION: Make sure that power is removed and all
capacitors have been discharged before measuring.
1) Set the rotary switch to "TEMP.
Ω
". OLis displayed.
2) Insert the black test lead to "COM" terminal and red test lead to
"
V -
Ω
-
µ
F
" terminal.
3) Short the test leads together and momentarily press the ZERO button to subtract
test lead resistance from measurement.
4) Touch the test leads to the circuit (Fuse Cartridge or other) and read resistance
value in the display.
5) Press
●
)
)
) button momentarily to enter continuity function if desired.
6) Repeat steps 3 and 4. The beeper sounds if continuity reading is less
than 10.0
Ω.
.
7) OL (overload) is displayed if the resistance across the input terminals is greater
than the full-scale rating on the range setting of the instrument. Be sure that the
contact between the probes and the circuit is clean. Dirt, oil, paint, rust or other for-
eign matter can seriously effect resistance measurements.
37
Содержание ACDC-620T
Страница 12: ...11 Figure 2 LCD Display ...
Страница 22: ...Figure 7 Display of Dynamic Recording 21 1 Press for more than 1 second to enter Dynamic Recording ...
Страница 31: ...Figure 17 Measuring AC Current CORRECT INCORRECT 30 ...
Страница 33: ...Figure 18 Measuring AC Curreft 32 ...
Страница 35: ...Figure 19 Measuring AC Motor Curent 34 ...
Страница 37: ...Figure 20 Measuring Voltage 36 ...
Страница 39: ...Figura 21 Measuring Resistance and Continuity FUSE CARTRIDGE 38 ...
Страница 41: ...Figure 22 Surface Temperature Measurement 40 Press Shift ...
Страница 43: ...Figure 23 Measuring Temperature on Refrigeration System Contact where you want to measuring PRESS T 42 ...
Страница 45: ...Figure 24 CAPACITANCEMEASUREMENTS 44 Negro Rojo ...
Страница 63: ...62 Figure 26 step 1 of Battery replacement Pull up slightly ...
Страница 64: ...63 Figure 27 Step 2 of Battery Replacement Pull and move to right ...