![AmpliVox PC850 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/amplivox/pc850/pc850_operating-manual_2935170043.webp)
42
D-0116493-B (OM015)
– PC850 Operating Manual
11.
USE WITH NON-MEDICAL ELECTRICAL EQUIPMENT
Any person who connects external equipment to signal input, signal output or other connectors has created a
medical electrical system and is therefore responsible for the system complying with the requirements of clause
16 of IEC 60601-1:2005
(General requirements for basic safety and essential performance).
If connections are made to standard equipment such as printers and computers, special precautions must be
taken in order to maintain medical safety. The following notes are provided for guidance in making such
connections to ensure that the general requirements of clause 16 of IEC 60601-1:2005 are met.
The following signal inputs and outputs on the Amplivox PC850 audiometer are electrically isolated to the
requirements of IEC 60601-1 in order to reduce any potential hazard associated with the use of mains-powered
equipment connected to these inputs and outputs:
Socket Label
Socket Type
Typical
Connection
USB
USB Connector
Type B
Computer
External equipment intended for connection to signal input, signal output or other connectors, shall comply
with the relevant IEC or international standards (e.g. IEC 60950, CISPR 22 & CISPR 24 for IT equipment, and
the IEC 60601 series for medical electrical equipment).
Equipment not complying with IEC 60601 shall be kept outside the patient environment, as defined in
IEC 60601-1:2005 (at least 1.5m from the patient). The operator must not touch the connected
equipment and the patient at the same time as this would result in an unacceptable hazard.
Refer to Diagrams 1 & 2 below for typical configurations of connected peripheral equipment.
Refer to Amplivox Ltd at the address given on the front of this user manual if advice is required regarding the