![AmpliVox PC850 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/amplivox/pc850/pc850_operating-manual_2935170030.webp)
29
D-0116493-B (OM015)
– PC850 Operating Manual
Click “ACCEPT” to exit the PC850 program and transfer the incomplete test thresholds to Audibase.
7.7.
INCONSISTENCY IN 1KHZ REPEAT TEST
If the repeat frequency option is enabled for a Computer test an additional test at 1kHz is carried out at the
end of the test and compared with the original test result at this frequency.
If the two results differ by 10dB or less then the second result is recorded as the threshold at 1kHz and the
test continues.
If the two results differ by more than 10dB the following dialogue box is displayed.
The operator can choose to stop the test (“Abort”), retry the repeat frequency (“Retry”) or skip the repeat
frequency and conti
nue the test (“Ignore”). If the Ignore option is selected the second result is recorded as the
threshold at 1kHz.
7.8.
USB LEAD DISCONNECTED
If the cable connecting the PC850 to the PC becomes disconnected any automatic test running will cease. It
may take the audiometer a few moments before the test stops. It may be possible to select control buttons and
menus options in the PC850 window on the PC but these will have no effect. Reconnecting the USB lead will
also have no effect.
Depending on the task in process when disconnection occurred, one or more error messages may be
displayed; for example:
If possible use the “File > Exit” drop-down menu selection at the top of the PC850 window. This should close
the PC850 window and return to Audibase. Further error messages may appear, for example: