SENDING THE RESULTS TO A PRINTER
27
D-0115692-C (OM031)
–
Otowave 202 Operating Manual
PLEASE WAIT
Trying to connect..
Do Not break IR link!
Cancel
The Otowave 202 will then attempt to connect to the printer. When this has been done the data will be
transferred. During this time the following message is displayed.
PLEASE WAIT
Sending Record
Do Not break IR link!
Cancel
If a connection cannot be made the print operation will time out (Able AP1300 printer only
– after approximately
30sec) and the following message will be displayed. The same message will occur if the print operation is
cancelled at this stage.
ERROR
Printing cancelled
No printer found
Next
Press
► to return to the PROCESS RECORD menu.
The infra-red link must not be broken once a connection has been established. If the printer or Otowave are
moved, or an object between them breaks the link, the printed results may be corrupted or the Otowave may
not respond to the controls until the printing process has timed-out (this could take 30 to 40 seconds). This
may also occur if the printer batteries are discharged while attempting to print.
Once the printing process has timed-out the resulting error message can be cleared and the results re-sent to
the printer. If the printed text is still corrupted select Cancel on the Otowave and then send the results to the
printer again.
For other error messages relating to printing refer to Section 13.
Note that, if required, it is possible to change to an alternative baseline display mode prior to printing. However
the baseline display mode that was stored in the instrument when the test was saved will always be retained.