IMPORTANT SAFETY INSTRUCTIONS
5
D-0115692-C (OM031)
–
Otowave 202 Operating Manual
2.3.
POWER SUPPLY OPTIONS
The tympanometer is designed for continuous operation and may be powered either by a mains adapter (which
is supplied, and specified as part of the equipment) or optional internal batteries.
Do not connect or disconnect the mains adapter lead while the instrument is operational
as this may cause it to shut down. Always switch off first (see Section 4.3).
Rechargeable batteries must be charged outside of the instrument
– they are not
charged by the mains adapter when this used.
Battery operation
Refer to Section 4.1 regarding the types of battery that may be used and their installation, replacement and
other precautions. Note that local regulations are likely to cover disposal of used batteries.
Mains operation
All other connections must be made before connecting the output lead from the adapter into the POWER
socket on the front face of the tympanometer. Switch on the mains supply - the indicator on the adapter will
illuminate green.
The output from mains adapter is fitted with electronic circuit protection. In case of overload the adapter will
shut down and the indicator will be extinguished. When the fault is cleared the adapter will operate as normal.
The input to the mains adapter is protected with a non-replaceable fuse. If this fails the adapter will not operate.
The mains adapter is the mains disconnect device and therefore the tympanometer should be positioned such
that easy access to the mains adapter is possible.
If a replacement mains adapter is required, please contact Amplivox or your Amplivox distributor.
2.4.
TYMPANOMETER CONNECTIONS
All the relevant accessory terminals and connections are labelled to ensure correct identification and
connection as follows:
Socket Label
Socket Type
Connected Part
Notes
PROBE
15-way D connector
Remote probe (electrical) *
AIR
4mm (nominal) Luer
Remote probe (pressure) *
CONTRA
3.5mm jack
Contralateral transducer *
USB Connector
Type B
Computer (via USB port)
See 2.6
POWER
2.5mm power jack
Mains AC/DC Adapter *
The relevant part numbers are indicated in Section 16.
For connected parts marked * only connect the accessories supplied with the instrument or supplied by
Amplivox or an Amplivox distributor. These parts have been tested for use with the Otowave 202 tympanometer
for compliance with the standards IEC 60601-1 and IEC 60601-1-2. The use of accessories other than those
specified may compromise compliance with these standards. For other sockets refer to Section 18.