7.3 NDIR CH
4
Electrical
Supply Voltage
10 - 16.5VDC
Max Power
Consumption
0.1W
Mechanical and Environment
Dimensions (H x W x D)
86.5mm x 85mm x 93mm
Weight
1kg
IP Rating
IP54
Humidity Range
15% to 90% r.h. non-condensing
Temperature Range
-20
˚C to + 40˚C
Pressure Range
±20kPa
Response Time (t90)
via diffusion
<30s
Warm-up Time
<5 mins (60 seconds typical depending on ambient temperature)
Stabilization time
(for calibration
adjustments)
3 mins
Sensor Unit
Detection Method
Non-dispersive infrared
Calibrated Range
0-5% v/v
Accuracy
±0.1% v/v or ±5% of indication
1
Drift
±0.1% v/v or ±5% of indication
1
over 3 months
Repeatability
±0.1% v/v or ±5% of indication
1
Sensing Element Life
>5 years
Resolution
0.01% v/v
2
Certification
IECEx
IECEx TRA 16.0020X
DPI
MDR 0001030 GD
1
Whichever is greater
2
The output measurement resolution is consistent across all gas types and does not infer the accuracy of
the gas type
Table 11: Typical NDIR Methane Cross-Sensitivity
Test Gas Mixture
Instrument variance from
Standard Test Gas
2% M 13% Oxygen in Nitrogen
±0.2% Methane
2% M 5% Carbon Dioxide in Air
±0.2% Methane
2% M 0.075% Ethane in Air
±0.2% Methane
2% M 0.020% Ethane in Air