Amos TEC 2602 Скачать руководство пользователя страница 7

 

 

TEC2602 Slide Drier

 

AMOSTEC2602OP20150317

 

---4--- 

period time and then keep quickly as 1 decrease. 

● 

5.2 Instructions 

1) Before  operation,  check  whether    every    part    of    the    machine    is    in    normal   

condition. 

2) Connect  the  machine  to  power  on  and  turn on the  power  switch  at  the  back  

cover.  Right  now  the  buzzer  sounds  and  the  LED display the ambient temperature,  

which  means  the  machine  is  in  normal  working  state. 

3)  Now  you  can  operate  the  machine   according  to  the  explanation  of  key 

instructions mentioned   above. 

4)Temperature error and regulation 
        The unit with high precision of temperature control already been adjust at factory before shipping, 

if customer feel the temperature control accuracy cann't meet demand, please calibration as the 

method: 

    For example: If  the setting temperature is 48

, After the unit is heating up thirty minutes about 

later, test the water temperature is Twater=49.3

by thermometer. 

        Calculation of the value necessary to compensate Tc=Tsetting-Twater

-1.3

℃)

 

     Press SET enter into the setting status, the green indicator lamp will light up, long press HEAT   

4 seconds more, LED display the current temperature compensation (-4

~+6

). If the 

temperature compensation displayed 2

 which request to plus the Tc(-1.3

℃)

to get the new 

compensate  Tc'

2

-1.3

=0.7

℃)

press 

 Or 

  to adust , and exit after five seconds 

automatically. 

● 

5.3 Function Declaration Of Alarm 

If  the  buzzer is sounds continuously,  that  means the machine malfunctioned : 

E01  After power on and heating, the temperature don't changes, the possible reasons is the digital 

sensor measurement failure or heating film, solid state relay is damaged or disconnected. 

E02   the digital sensor measures the temperature is over 100 degree, the solid state relay or sensor 

meet problem. 

E03   the digital sensor measures the temperature is under -10 degree, please preclude the ambient 

temperature reason firstly. If the ambient temperature within the requirement, the possible reason is 

the digital sensor is damaged, no signal input, please replace the new sensor. 

 

Содержание TEC 2602

Страница 1: ...TEC 2602 Version AMOSTEC2602OP20150317 Slide Dryer Operation Manual www amos scientific com AMOS SCIENTIFIC PTY LTD ABN37 159 778 140...

Страница 2: ...nd the matters needs attention for safety reasons Please carefully read this manual before using so that you can understand its functions better and make full use of the products If you have any quest...

Страница 3: ...aracteristics apply condition 2 3 Main structures Of Slide Dryer 2 4 Technical Data 3 5 Operation Manual 3 5 1Explaination Of Functional Keys 3 5 2Instructions 4 5 3Function Declaration Of Alarm 4 6 L...

Страница 4: ...es The equipment should be installed far away from any interference source Caution Hot Electrical Protection Ratings I class B type Shell protection grade IP53 Baleful liquid Leak in proof degree Norm...

Страница 5: ...Make sure that the product and accessories do not get damaged or missing during transportation 2 Functions characteristics apply condition The Slide Dryer is made of sheet aluminum and has the advant...

Страница 6: ...s SET to exit the setting status Or the system will test and exit the setting status automatically if SET not pressed to exit HEAT Press this key to select HEAT ON or HEAT OFF the machine is on the he...

Страница 7: ...Calculation of the value necessary to compensate Tc Tsetting Twater 1 3 Press SET enter into the setting status the green indicator lamp will light up long press HEAT 4 seconds more LED display the c...

Страница 8: ...TEC2602 Slide Drier AMOSTEC2602OP20150317 5 6 Layout Of Components Wire Diagram Of Electric Equipment...

Страница 9: ...05 34 P06 33 P07 32 P20 21 P21 22 P22 23 P23 24 P24 25 P25 26 P26 27 P27 28 PSEN 29 ALE P 30 TXD 11 RXD 10 VCC 40 GND 20 5V C3 89S52 5V 1 2 3 4 SSR1 5V 5V 5V C1 C2 S1 S2 S3 S4 R19 R20 R18 R17 5V 1 2 3...

Страница 10: ...the receptacula as figure 3 shown Before replacing the fuse must turn off the power switch and disconnect the power Specification of fuse 3A 220V 5A 110V 9 Trouble shooting Symptom Cause Judgment Reme...

Страница 11: ...Standard Accessories List No Accessory Name Qty Notes 1 Slide Dryer 1 unit 2 Power Cord 1pc 3 Fuse 2pcs 4 Operation Manual 1pc...

Страница 12: ...Published by AMOS SCIENTIFIC PTY LTD 7 Gabrielle Court Bayswater North VIC 3153 Melbourne Australia Tel 61 410190188 Fax 61 410190188 ABN 37 159 778 140 www amos scientific com...

Отзывы: