TEC2602 Slide Drier
AMOSTEC2602OP20150317
---7---
8. Cleaning
&
Maintenance
●
Prior to operation, please earnestly read this Operation Manual and carry out by this
Manual.
●
Keep the instrument cleaning. Please clean up the remaining substance and dirt
after work finished.
●
Forbid using sharp hard object to harm the worktable.
●
Install fuse as figure(1) shown then put it into the installation part as figure(2)shown.At
last put the whole installation into the receptacula as figure(3)shown.
●
Before replacing the fuse must turn off the power switch and disconnect the
power.Specification of fuse: 3A
(
220V
)
,5A(110V).
9
.
Trouble shooting
Symptom
Cause
Judgment
Remedy
No display
Poor connection of power
supply
Connection again
Connection well
Switch at the rear of the
equipment not on
Turn on
Turn on
Fuse blow
Take out the fuse to check its
internal resistance by
multimeter, if the value is
infinite, the fuse is broken
Replace a new one
Buzzer sounds
Temperature sensor was
damaged
Connect a new sensor to
terminal to start machine, if
all be okay, it means the
sensor is broken
Replace a new one
2
3
1