background image

 

 

ASD 220 Slide Dryer   

Service Manual 

---3--- 

   

D

the instrument is not heating after power on . 

    1

):

heating film is damaged 

            Maintain 

use the Philip screwdriver to loosen the six screws in the base-board, 

take down the base-board and you will see the heating film as it 

showed  in  the  picture.  And  then  loosen  the  four  screws  on  it  to 

replace the heating film. 

    2

):

Solid state relay is broken 

          Maintain

use the Philip screwdriver to loosen the six screws in the base-board, 

take down the base-board and you will see the solid state relay as it 

showed  in  the  picture.  Take  down  the  damaged  SSR  and  replace  a 

new SSR. 

 

 

 

 

 

 

 

 

 

 

Содержание ASD220

Страница 1: ...ASD 220 Slide Dryer Service Manual www amos scientific com AMOS SCIENTIFIC PTY LTD ABN37 159 778 140 ...

Страница 2: ......

Страница 3: ... please connect the power supply again or replace the fuse 2 power module is damaged Maintenance Use Philip screwdriver to loosen the six screw in the base plate and take down the base plate and there will be a power source module on the main board in the instrument as it shows on the picture And take down the power source module from the main board by hand if it is difficult to take down the powe...

Страница 4: ...will see the digital sensor as it showed in the picture Loosen the screw in the sensor and then replace the sensor 2 IC is broken Maintain use the screwdriver to loosen the screw and you will see the IC as it showed in the picture you can use the flat apparatus as flat tip screw driver to take down the IC by inserting the flat apparatus in the slot and tipping it up slightly and then take it down ...

Страница 5: ...down the base board and you will see the heating film as it showed in the picture And then loosen the four screws on it to replace the heating film 2 Solid state relay is broken Maintain use the Philip screwdriver to loosen the six screws in the base board take down the base board and you will see the solid state relay as it showed in the picture Take down the damaged SSR and replace a new SSR ...

Страница 6: ...ASD 220 Slide Dryer Service Manual 4 4 Wiring Diagram ...

Страница 7: ...ASD 220 Slide Dryer Service Manual 5 5 Schematic Diagram ...

Страница 8: ...Published by AMOS SCIENTIFIC PTY LTD 37 632 CLAYTON ROAD CLAYTON SOUTH VIC 3169 AUSTRALIA Tel 61 3 95518850 Fax 61 3 95516885 ABN 37 159 778 140 www amos scientific com ...