Appendix A – Chapter 2 Mobile Patient Bed Operation
Page 258 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
1. Place the bed in the lowest position.
2. Place the control pendant in the holster.
3. Ensure that the AC power cord is disconnected from the wall outlet, if applicable.
4. Unlock the casters, as described in
5. If desired, activate directional lock mode, as described in
on page
.
6. Move the bed to the desired location.
CAUTION
Do not push or pull the Mobile Patient Bed using the headrest or headrest assembly. Applying
excessive pressure to the headrest or headrest assembly may damage these components.
2.5 Positioning the Bed Next to the
CATALYS®
System
The Mobile Patient Bed should be placed next to the
CATALYS®
System to accommodate the
following three bed positions:
• Patient load position–The bed is rotated approximately 50° out from under the
CATALYS®
System. This position allows for patient loading and unloading and provides clearance for
the bed to be in the upright seated or flat position.
•
LIQUID OPTICS™
Interface installation position–The bed is rotated approximately 19° out
from under the
CATALYS®
System. This position allows the physician convenient access to
the treatment eye for
LIQUID OPTICS™
Interface installation.
• Patient treatment position–The bed is rotated to the 0° position under the
CATALYS®
System with the patient treatment eye located directly under the laser aperture. This
position allows the physician to dock the patient to the
CATALYS®
System and perform
Cataract Laser Surgery.
2.6 Connecting the Bed to the
CATALYS®
System
Refer to
Connecting the Mobile Patient Bed to the CATALYS® System
on connecting the bed to the
CATALYS®
System.
The
(connect) button
on the control pendant indicates the connection status, as follows:
• Slowly blinking blue—Mobile Patient Bed is attempting to connect to
CATALYS®
System
• Solid blue—Mobile Patient Bed is connected to
CATALYS®
System
• Rapidly blinking blue—Mobile Patient Bed has been disconnected from
CATALYS®
System
during procedure
The
(Bluetooth connection status) icon on the
CATALYS®
System touchscreen also indicates
the connection status, as follows:
• Solid blue—Mobile Patient Bed is connected to
CATALYS®
System
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...