Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 181 of 314
Software Release: cOS 5.00
NOTE
The following three sections pertain only to the Mobile Patient Bed. If using the patient chair,
proceed to
.
Connecting the Mobile Patient Bed to the CATALYS® System
After verifying the information on the Surgical Timeout Screen and pressing the APPROVE
button, the Docking Screen guides you through the patient docking procedure. When the initial
Docking Screen displays, the Connection panel is open. Establish Bluetooth wireless connection
between the Mobile Patient Bed and the
CATALYS®
System, as follows:
1. Line up the Mobile Patient Bed with the
CATALYS®
on page
for instructions). The maximum recommended
distance between the bed and the
CATALYS®
System is 10 ft (3 m) when establishing
Bluetooth connection.
2. Press the
(connect) button
on the Mobile Patient Bed control pendant to establish
Bluetooth connection between the Mobile Patient Bed and the
CATALYS®
System. The
button
blinks slowly to indicate that the Mobile Patient Bed is attempting to connect to
the
CATALYS®
System.
3. When the Bluetooth connection is successful, the
(connect) button on the control
pendant illuminates blue; the
(Bluetooth connection status) icon at the top of the
CATALYS®
System screen turns from gray to blue; and the Mobile Patient Bed name
displays to the right of the icon, as shown in the following figure.
Figure 4.21 Bluetooth Connection Status Icon and Mobile Patient Bed Name
Mobile Patient Bed Name
Bluetooth Connection Status Icon
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...