3 Instrument Maintenance
Periodic preventative maintenance will prolong the life of the instrument. The following steps are
recommended:
● If the instrument is dirty or oily, allow it to soak in warm, soapy water before cleaning with
a rag or soft brush. When finished, rinse with fresh water to remove any residual soap or
dirt.
● Before each use:
○ Check for proper installation of the X2change™ sensor (Appendix B)
○ Check for nicks and cuts on the cable.
● After each use:
○ Clean and rinse the instrument using fresh water.
○ Dry the instrument completely, and store it in a cool, dry place.
● Monthly:
○ Apply a layer of Molykote 44 medium silicone grease to the female half of the
connection as shown and fully insert the plug. Wipe away any excess that
squeezes out. This will lubricate both male and female connectors.
● Yearly:
○ Send the instrument or X2change
TM
sensors to a service centre for diagnostics
and re-calibration
AML Oceanographic Ltd.
12/18