As required by NFPA-99 (National Fire Protection Association), Health Care Facilities oxygen systems having the source 

of supply outside the building shall have an inlet to connect a temporary, auxiliary source of supply for emergency 

or maintenance situations. This connection shall include necessary valves and to allow for an emergency supply 

of oxygen and isolation of the normal source of supply. The inlet connection shall be a 1” female NPT connection 

and be provided with a plug to prevent entry of foreign matter into the oxygen system when not in use.

Amico Emergency Oxygen Inlet Stations will perform in conformity with the  descriptions contained in this manual 

and accompanying labels and/or inserts when  assembled,  operated, maintained and repaired in accordance with the 

instructions provided.

Emergency Oxygen Inlet Stations must be inspected periodically. A defective product should not be used. Parts that 

are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should such repairs or 

replacements become necessary, please contact your local distributor.

Introduction

Medical Facility Responsibility

Содержание M-FILL-OXY-LP

Страница 1: ...Operation Maintenance Installation Manual Emergency Oxygen Inlet Station Low Pressure 1 Recessed Mount Model M FILL OXY LP Surface Mount Model M FILL OXY LP S w w w a m i c o c o m...

Страница 2: ...edical Facility Responsibility 3 Installation of Emergency Oxygen Inlet Station 4 Operating Instructions 4 Repair and Replacement Parts List 5 Schematic Drawings 6 7 Optional Check and Relief Valves P...

Страница 3: ...rovided with a plug to prevent entry of foreign matter into the oxygen system when not in use Amico Emergency Oxygen Inlet Stations will perform in conformity with the descriptions contained in this m...

Страница 4: ...em for leaks as per appropriate standards To Operate 1 Ensure that the isolation valve is closed 2 Open the threaded connection 3 Connect a pigtail to the oxygen threaded connection 4 Check the inlet...

Страница 5: ...lve seats and O rings 2 V X VLV HAND M Valve handle sleeve for 1 3 V X GAUGE 100 Gauge 100 psi 690 kPa 1 8 NPT 4 M X OXF PLUG LP 1 NPT hex head plug 5 V X BOX PLUG Box plug for oxygen inlet station Bo...

Страница 6: ...Connect to pipeline Inlet Oxygen connection 1 NPT Shutoff Valve Provision for Padlock 7 1 4 184 Main Line Shutoff Valve Check Valve 2 required Relief Valve 75 PSIG Turn Down and Screen Main Oxygen Sup...

Страница 7: ...XY LP2 Emergency low pressure oxygen inlet station 2 recessed mounted M FILL OXY S Emergency high pressure oxygen inlet station 1 surface mounted M FILL OXY Emergency high pressure oxygen inlet statio...

Страница 8: ...co com 85 Fulton Way Richmond Hill Ontario L4B 2N4 Canada Toll Free Tel 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email info amico com APE OM INSTAL EOIS LP 2IN 01...

Отзывы: