14
Amico Corporation
REPEAT ALARM
Factory Default - Disable
Disable
Set switch #1 to the OFF position.
Set switch #2 to the OFF position
Enable
1 HR
Set switch #1 to the ON position.
Set switch #2 to the OFF position
Enable
12 HR
Set switch #1 to the OFF position.
Set switch #2 to the ON position
Enable
24 HR
Set switch #1 to the ON position.
Set switch #2 to the ON position.
NOTE:
The above repeat alarm only applies to the Master/NEMA 4 alarm.
Chart of Gas Specific Settings of Dip-Switches
Master/NEMA 4 Status Module
A l e r t - 2 S e r i e s
P a g e : 2 1
SETTING GAS IDENTIFICATION SWITCHES
NOTE: DO NOT TAMPER WITH SWITCHES #1, 2 AND 3
ON THE DIP-SWITCH. TAMPERING WITH THESE
POSITIONS WILL RESULT IN AN ERROR MESSAGE
BEING DISPLAYED (EO2) AND WILL DISABLE THE
ELECTRICAL INTERLOCK FROM THE GAS SPECIFIC
SENSOR.
CHANGES TO THESE SWITCHES SHOULD ONLY BE
DONE BY PROPERLY TRAINED PERSONNEL, WHEN
CIRCUIT BOARDS HAVE TO BE CHANGED IN THE FIELD.
Switches # 1, 2 and 3 are used for the gas identification of the
display module. These will be set at the factory and should not
be tampered with in the field.
CHART OF GAS SPECIFIC SETTINGS OF DIP-SWITCHES
O
N
O
N
O
N
O
N
O
N
O
N
O
N
O
N
4
4
4
4
4
4
4
4
3
3
3
3
3
3
3
3
2
2
2
2
2
2
2
2
1
1
1
1
1
1
1
1
5
5
5
5
5
5
5
5
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
9
9
9
9
9
9
9
9
10
10
10
10
10
10
10
10
#1 - off
#2 - off
#3 - on
#1 - off
#2 - off
#3 - off
#1 - off
#2 - on
#3 - off
#1 - on
#2 - on
#3 - on
#1 - on
#2 - off
#3 - on
#1 - on
#2 - on
#3 - off
#1 - off
#2 - on
#3 - on
#1 - off
#2 - off
#3 - off
Oxygen
Nitrogen
Vacuum
Carbon Dioxide
Medical Air
WAGD/AGSS
Nitrous Oxide
Instrument Air
1 HR
12
HR
24
HR