background image

     www.amico.com     

3

Intended Use

Introduction

The AC 200 bed is intended for patients in all areas of the hospital and in free-standing surgical centers.

The AC 200 bed is intended for patients in the medical surgical units of hospitals.

The AC 200 features height adjustment, a reclining back rest and regulating system for leg positioning, with a central 

locking system. 

The motions of the AC 200 are achieved by independent DC electric motors through controls located on the siderails of 

the bed allowing positions to be changed. The motors are equipped with overheat protection. 

The bed is also equipped with a backup battery to allow caregivers to adjust positions during patient transport and 

unexpected power outages. 

This Manual provides instructions required for normal operation of the AC 200 bed from Amico Beds Corporation. Prior 

to operation of this bed, it is important that this manual is read and ensure that all safety aspects contained in this 

manual are strictly adhered to.

Содержание AC-200C

Страница 1: ...Operating Maintenance Manual AC 200C...

Страница 2: ...e Head Foot Board 6 Siderails 6 Lowering the Siderails 6 Raising the Siderails 6 Central Locking System 7 Locking the Bed 7 Unlocking the Bed 7 Using the CPR Release Lever 7 Electrical 7 Standard Feat...

Страница 3: ...are achieved by independent DC electric motors through controls located on the siderails of the bed allowing positions to be changed The motors are equipped with overheat protection The bed is also e...

Страница 4: ...manual The bed has to be operated by trained personnel only Maximum weight limit is 550 lbs 250 kg Entering and exiting the bed has to be done at the side with the siderails lowered completely Avoid...

Страница 5: ...gle G Auto Contour Up Press on this button to lower the fowler and knee gatch section to the desired position H Auto Contour Down Press on this button to raise the fowler and knee gatch section to the...

Страница 6: ...headboard or footboard Once both of the mechanisms are up they will hold in place Grab the sides of the head foot board with both hands tilt outwards and pull up Siderails LOWERING THE SIDERAILS To lo...

Страница 7: ...ush the bed down with the other hand when the bed is in required position let go of the lever It s recommended that the fowler section is completely lowered to a flat position when the CPR handle is u...

Страница 8: ...he siderails to the control box vice versa Verify electrical connections from the foot board to the control box vice versa Measure voltage level of batteries Battery voltage is measured correct bed ba...

Страница 9: ...s originally designed Pay particular attention to safety features including but not limited to the following Siderail latching mechanisms Caster braking systems Integrity of sleep surface ticking Inte...

Страница 10: ...h warm water containing soap or mild detergent To rinse After washing rinse cover with warm water to remove residue To dry Thoroughly wipe with a dry sponge or soft cloth To disinfect Wash the surface...

Страница 11: ...s three 3 years from the date of shipment Within this period Amico Beds Corporation will replace any part which is proven to be defective at no charge Shipping and Installation costs after the first t...

Страница 12: ...amico com Amico Beds Corporation 85 Fulton Way Richmond Hill ON L4B 2N4 Canada Toll Free Tel 1 877 462 6426 Tel 905 764 0800 Fax 905 764 0862 Email info amico com www amico com TTS OM AC200C 10 23 20...

Отзывы: