For any other issue not covered in this section, contact AMI
at 714.848.5533 or visit
us at www.
amio2
.com
for support.
END OF TROUBLE SHOOTING, MAINTENANCE & REPAIRS
Sealing/Ingress Protection Maintenance
Whenever the Adalet Explosion-proof cap is opened, visually inspect the O-ring for any signs of
damage or excessive wear.
Action
:
•
If the O-ring needs to be replaced, contact AMI
IMPORTANT MESSAGE ABOUT REPAIRS
Where repair is possible:
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
LE REMPLACEMENT DE COMPOSANTS PEUT COMPROMETTRE LA SECURITE INTRINSEQUE.
WARNING
IMPORTANT MESSAGE ABOUT CLEANING REQUIREMENTS
The Analyzer is designed to function properly without cleaning requirements.
Sensor Replacement
It is recommended that the sensor be replaced when the Span Factor exceeds a value of 980.
Action
:
•
Refer to pages 20 – 21 for instructions on how to replace the sensor
•
Refer to page 23 for instructions on how to view the Span Factor
Analyzer Calibration
For the best accuracy, it is recommended that the Analyzer is calibrated every 30 to 45 days.
Action
:
•
Refer to pages 22 to 23 for instructions on how to perform a calibration
MAINTENANCE