M O D E L 2 0 5
O P E R A T O R ’ S M A N U A L
48
48
48
48
Note
To prevent unauthorized personnel from accessing the calibration feature a temporary
one-day password is required. Contact the AMI Service department for this password.
You will need the Power Supply serial number displayed on the HOME screen to obtain
this password.
5. Input the one-day password and press ENTER.
6. Setup the Digital Multimeter to read the 200 mVDC scale.
7. Prepare to insert the test probes into the red and black terminals on the shunt.
Any testing or recording instruments connected to any points in the machine or to the
welding head may be exposed to extremely high transient voltage at the time the arc starter
operates. ENSURE THAT ALL TEST EQUIPMENT IS DISCONNECTED DURING ARC START.
Make sure that when an arc is struck nothing will be overheated or melted.
Note
If an orbital weldhead is used instead of a manual torch you will need to hook up the
weldhead arc gas line to the power supply