2-4
84-55
TROUBLESHOOTING
9.2.2 Air Blower fails to
start
Blower power on switch
is off
No power to Blower
motor
Turn the switch on
(SSW2)
Check Fuse F1 if blown
then change, still not
power on Check Blower
Motor supply terminals
for power 220 V AC &
rectify by the help of
wiring diagram if wiring
ok then Blower motor is
faulty contact AMF
Reece for new & replace
9.2.3 All Sensors & MMI
fails Power on
24 V Power supply unit
not powered
Check Fuse F2 if Blown
then Change if still no
Green LED at Power
Supply Unit then Power
Supply faulty contact
AMF Reece for new &
replace
9.2.4 Table Lamp fails to
on
Power on switch off
Lamp Bulb Blown
Lamp unit is faulty
Turn on power switch at
two locations one at lamp
head & another at Lamp
Base
Change the bulb
Rectify by the help of
wiring diagram if wiring
ok then Lamp Unit faulty
contact AMF Reece for
new & replace
9.2.5 Control Cabinet
Cooling Fan Not
Functional
Fan not powered
Fan is faulty
Check Fuse F3 if blown
then change
Check power supply 24 V
DC at Fan Input
terminals, rectify by the
help of wiring diagram If
still not working then fan
is faulty contact AMF
Reece for new & replace