3-18 | TM 2000 Oxygen Analyzer
Sensor Connections
Follow all general grounding and shielding requirements as de
-
scribed in the “General Wiring Requirements” section. Also make
sure that AC mains power has been removed from the control unit
and sensor before making sensor connections.
Follow the connection instructions for the applicable options for your analyzer. All
sensor connections require accessing the sensor board (Figure 3-19).
Sensor cabling must be twisted pairs in rigid metal conduit or in an
overall braided shield. Follow wiring guidelines as specified in Figure
3-17.
Sensor AC mains supply connections
Connections for AC mains supply to the sensor are labeled as follows:
L
Line Connection
L2
Neutral (USA)
Chassis Stud
Equipment Ground (Protective Conductor)
NOTE
Figure 3-19.
Sensor connections.
Содержание TM2000 Pro
Страница 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 33: ...Installation and Start Up 3 11 Figure 3 10 Weatherproof panel mount Figure 3 11 Weatherproof wall pole mount...
Страница 35: ...Installation and Start Up 3 13 Figure 3 14 General purpose rack mount Figure 3 15 General purpose wall mount...
Страница 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...