13.2 Coupling possibilities IEC/EN 61000-4-5 / ANSI C62.45
EUT input connections
Combination wave pulse
IEC/EN 61000-4-5 / ANSI C62.45:
1-phase
L1, N, PE
2-phase
L1, L2, N, PE
3-phase
L1, L2, L3, N, PE
Line to ground (12
Ω)
up to 15 kV L1
PE
L1
PE
L1
PE
Line to ground (2
Ω)
up to 8 kV N
PE
L2
PE
L2
PE
N
PE
L3
PE
N
PE
Lines to ground (12
Ω) up to 15 kV L1, N
PE
L1, L2
PE
L1, L2
PE
Line to ground (2
Ω)
up to 8 kV
L1, N
PE
L1, L3
PE
L2, N
PE
L1, N
PE
L1, L2, N
PE L2, L3
PE
L2, N
PE
L3, N
PE
L1, L2, L3
PE
L1, L2, N
PE
L1, L3, N
PE
L2, L3, N
PE
Line to line (2
Ω & 12 Ω) up to 15 kV
L1
N
L1
L2
L1
L2
L1
N
L1
L3
L2,
N
L1
N
L2
L1
L2
L3
L2
N
L3
L1
L3
L2
L3
N
Line to line (12
Ω)
up to 15 kV
L1, L2
N
L1, L2
N
Line to line (2
Ω)
up to 8 kV
L1, L3
N
L2, L3
N
L1, L2, L3
N
All ANSI C62.45 couplings with 2
Ω
up to 8 kV
All ANSI C62.45 couplings with 12
Ω up to 15 kV