Operation
SPECTRO MIDEX — 31.08.2016
— 21
5.5 Standby
mode
Comply with the following instructions regarding the standby mode:
If the instrument has been down for an extensive period (weekends, vacation
shutdown), the X-ray tube must be run under standby conditions for at least
20 minutes. Start measuring only after this period.
If the instrument has been down for an extensive period ( > 1 month), the X-
ray tube must be run under standby conditions for at least 1 hour. Start
measuring only after this period.
To activate/deactivate the standby mode, click on the
symbol on the task
bar at bottom right. This symbol can be used to either “activate” or
“deactivate” the standby mode.
Recalibrate the instrument (see Chapter 6.2).
or
Measure a reference sample.
5.6
Carrying out the measurement
Caution! Risk of crushing injuries!
When closing the sample chamber lid, there is an inherent risk of crushing injuries.
When closing the sample chamber lid keep the fingers out of the sample
chamber.
Caution! Risk of injury due to laser radiation!
The instrument uses a laser pointer of protection class II. Your eyes will suffer
damage if you look at this laser pointer for an extended period of time.
Therefore never look directly at the laser pointer.