Remote
Operation
5-58
M370430-01 Rev F
SCPI Preset Status
Configures the status data structures to ensure that certain events are
reported at a higher level through the status-reporting mechanism.
These events are summarized in the mandatory structures, the Operation
Status Register and Questionable Status Register.
The PRESet command affects only the enable registers and the
transition filter registers of the status data structures. PRESet does not
clear any of the event registers or any item from the error/event queue.
The *CLS command is used to clear all event registers and queues in the
device status-reporting mechanism.
For the device-dependent status data structures, the PRESet command
sets the enable register to all 1s and the transition filter register to report
only positive transitions. For the SCPI mandatory status data structures,
the PRESet command sets the transition filter registers to recognize only
positive transitions and sets the enable register to 0s. The following will
not be affected by this command: Service Request Enable Register,
Parallel Poll Enable Register, the memory register associated with the *
SAV command, the power supply address, Output Queue, and the
power-on-status-clear flag setting.
Table 5-19
Preset Values of User Configurable Registers
Register
Filter/Enable
Preset Value
Operational
Operational Enable Register
0
Positive Transition Filter
1
Negative Transition Filter
0
Questionable
Questionable Enable Register
0
Positive Transition Filter
1
Negative Transition Filter
0
All others
All other Enable Register
1
Positive Transition Filter
1
Negative Transition Filter
1
SCPI command:
[:]STATus[<channel>]:PRESet
Содержание Sorensen XG 100-15
Страница 2: ......
Страница 21: ...1 Introduction Chapter 1 Introduction describes the features of the XG Family Programmable DC Power Supply ...
Страница 93: ...Local Operation M370430 01 Rev F 3 47 3 Figure 3 5 Series Operation ...
Страница 99: ...Local Operation M370430 01 Rev F 3 53 3 ...
Страница 100: ...Local Operation 3 54 M370430 01 Rev F ...
Страница 220: ...Communications Options 6 14 M370430 01 Rev F ...
Страница 224: ...Communications Options 6 18 M370430 01 Rev F 9 In the Setup Type window choose Typical and click Next ...
Страница 225: ...Communications Options M370430 01 Rev F 6 19 6 10 Check that your Settings are correct and then click Install ...
Страница 232: ...Communications Options 6 26 M370430 01 Rev F 9 In the Setup Type window choose Typical and click Next ...
Страница 244: ...Communications Options 6 38 M370430 01 Rev F ...
Страница 252: ...Communications Options 6 46 M370430 01 Rev F ...
Страница 292: ......
Страница 329: ...M370430 01 Rev F A 37 A APPENDIX A SCPI Command Reference ...
Страница 330: ......
Страница 348: ...APPENDIX C Specifications C 12 M370430 01 Rev F Figure C 2 XG 850 Watt Mechanical Dimensions 60 V to 600 V Models ...
Страница 349: ...APPENDIX C Specifications M370430 01 Rev F C 13 Figure C 3 XG in Full Rack Mechanical Dimensions 6 V to 40 V Models ...
Страница 351: ...APPENDIX C Specifications M370430 01 Rev F C 15 Figure C 5 XG in Full Rack Mechanical Dimensions 60 V to 600 V Models ...
Страница 354: ...APPENDIX C Specifications C 18 M370430 01 Rev F Figure C 9 XG850 Low Voltage Model Dimensions Top Side and Rear ...
Страница 361: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 5 Figure D 3 RM XG1 Rack Rails ...
Страница 367: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 11 Figure D 10 Installing Chassis ...
Страница 368: ...APPENDIX D Rack Mount Options and Installation Instructions D 12 M370430 01 Rev F ...
Страница 369: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 13 ...
Страница 370: ......
Страница 395: ...Rack Mount Options and Installation Instructions M370430 01 Rev F E 25 ...
Страница 396: ......
Страница 399: ......