Installation and Configuration
Remote/Local Operation
22
Operating Manual for GPIB for XHR/XFR Series Power Supply
If the rear panel PON REM switch is set to 1, the power supply will power up in
local mode. Power supply control is at the front panel. During an operating
session, you can toggle between local mode and remote mode by using the front
panel LOCAL button and sending commands as described in “Remote Mode
Operation” on page 22 and “Local Mode Operation” on page 23.
Table 2.5
Remote Mode Power On Conditions
Remote Mode
Operation
While in remote mode, use the interface commands to control the output of the
power supply from a computer. See Section 3, “Operation” for a complete list of
device-dependent commands available with this interface. For information about
how to use GPIB commands and GPIB lines, refer to the documentation for the
particular GPIB controller interface installed on your computer.
Condition
Default Settings
7.5-140 Model Example
Voltage
0 V
VSET 0
Current
0 A
ISET 0
Soft Voltage Limit
VMAX (see models)
VMAX 7.5
Soft Current Limit
IMAX (see models)
IMAX 140
OVP Trip Voltage
Model VMAX + 10%
OVSET 8.25
Delay
0.5 s
DLY 0.5S
Foldback Protection
OFF
FOLD OFF
Output
ON
OUT ON
Hold
OFF
HOLD OFF
Unmask
NONE
UNMASK NONE
Service Request Capability
OFF
SRQ OFF
AUXA
OFF
AUXA OFF
AUXB
OFF
AUXB OFF