Features and Specifications
Specifications
13
Table 1.5
Specifications for XHR 1000 W Series Supply with GPIB Interface Installed (7.5 V to 60 V)
Models
7.5-130
20-50
33-33
40-25
60-18
Program Resolution
Voltage
Current
OVP
1.16mV
42.0mA
1.16mV
1.8mV
30.8mA
1.8mV
3.08mV
18.2mA
3.08mV
6.2mV
9.8mA
6.2mV
9.2mV
6.44mA
9.2mV
Program Accuracy
1
Voltage
Current
OVP
1.
Apply accuracy specifications according to the following voltage program accuracy example:
Set a model 20-50 power supply to 10 volts.
The expected result will be within the range of 10 volts ± 50mV ± 0.12% of the set voltage of 10 volts.
10mV
±0.12%
900mA
±0.1%
80mV
50mV
±0.12%
750mA
±0.1%
150mV
75mV
±0.12%
500mA
±0.1%
200mV
75mV
±0.3%
350mA
±0.1%
400mV
150mV
±0.25%
250mA
±0.1%
600mV
Readback Resolution
Voltage
Current
1.16mV
42.0mA
1.8mV
30.8mA
3.08mV
18.2mA
6.2mV
9.8mA
9.2mV
6.44mA
Readback Accuracy
1
Voltage
Current
30mV
±0.12%
900mA
±0.1%
60mV
±0.12%
750mA
±0.1%
75mV
±0.12%
500mA
±0.1%
75mV
±0.3%
350mA
±0.1%
150mV
±0.25%
250mA
±0.1%
Содержание Sorensen GPIB-XFR
Страница 2: ......
Страница 4: ...This page intentionally left blank ii...
Страница 6: ...This page intentionally left blank iv...
Страница 8: ...This page intentionally left blank vi...