Sorensen DCS Series 1kW and 1.2kW Supplies
Features and Specifications
M362500-01 Rev K
1-3
1.4
Specifications
1.4.1
Electrical Specifications for 1kW Models
Specifications are warranted over a temperature range of 0-50°C with default local sensing.
From temperatures of 50-70°C, derate output 2% per °C.
Models
8-125 10-100 20-50
33-33
40-25 50-20 60-18 80-13 100-10 150-7 300-3.5 600-1.7
Output Ratings:
Output Voltage
Output Current
Output Power
0-8V
0-125A
1000W
0-10V
0-100A
1000W
0-20V
0-50A
1000W
0-33V
0-33A
1089W
0-40V
0-25A
1000W
0-50V
0-20A
1000W
0-60V
0-18A
1080W
0-80V
0-13A
1040W
0-100V
0-10A
1000W
0-150V
0-7A
1050W
0-300V
0-3.5A
1050W
0-600V
0-1.7A
1020W
Line Regulation:
1
Voltage
Current
8 mV
125 mA
10 mV
100 mA
20 mV
50 mA
33 mV
33 mA
40 mV
25 mA
50 mV
20 mA
60 mV
18 mA
80 mV
13 mA
100
mV
10 mA
150 mV
7 mA
300 mV
3.5 mA
600 mV
1.7 mA
Load Regulation:
2
Voltage
Current
8 mV
125 mA
10 mV
100 mA
20 mV
50 mA
33 mV
33 mA
40 mV
25 mA
50 mV
20 mA
60 mV
18 mA
80 mV
13 mA
100
mV
10 mA
150 mV
7 mA
300 mV
3.5 mA
600mV
1.7 mA
Meter Accuracy:
Voltage
Current
0.09V
1.35A
0.11V
1.1A
0.3V
0.6A
0.43V
0.43A
0.5V
0.35A
0.6V
0.30A
0.7V
0.28A
0.9V
0.23A
1.1V
0.11A
1.6V
0.08A
4.0V
0.045A
7.0V
0.018A
OVP Adjustment Range 0.4-8.8V 0.5-11V 1.0-22V 1.65-36.3V
2-44V
2.5-55V
3-66V 4-88V 5-110V
7.5-165V 15-330V 30-660V
Output Noise/Ripple(V)
3
Rms
p-p
(20Hz-20MHz)
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
4 mV
60 mV
6 mV
60 mV
12 mV
160 mV
20 mV
200 mV
50
mV
300
mV
Stability:
4
Voltage
Current
4 mV
62.5 mA
5 mV
50 mA
10 mV
25 mA
16.5 mV
16.5 mA
20 mV
12.5 mA
25 mV
10 mV
30 mV
9 mA
40 mV
6.5 mA
50 mV
5 mA
75 mV
3.5 mA
150 mV
1.75 mA
300 mV
0.85 mA
Temperature
Coefficient:
5
Voltage
Current
1.6 mV
37.5 mA
2 mV
30 mA
4 mV
15 mA
6.6 mV
9.9 mA
8 mV
7.5 mA
10 mV
6 mA
12 mV
5.4 mA
16 mV
3.9 mA
20 mV
3 mA
30 mV
2.1 mA
60 mV
1.05 mA
120 mV
0.51 mA
Maximum Remote
Sense Line Drop
6
Compensation/line
4V
4V
4V
4V
4V
4V
4V
4V
4V
4V
4V
4V
1
For input voltage variation over the AC input voltage range, with constant rated load
2
For 0-100% load variation, with constant nominal line voltage
3
Typical P-P noise and ripple
4
Maximum drift over 8 hours with constant line, load, and temperature, after 20-minute warm-up
5
Change in output per °C change in ambient temperature, with constant line and load
6
Line drop subtracts from the maximum available output voltage at full rated power
AC Input:
200-250 Vac at 9A rms or 100-130 Vac at 18A rms, 47-63 Hz
Maximum Voltage Differential from output to safety ground:
150 Vdc