Sorensen DCS-E Series 3kW Supplies
Theory of Operation
M362295-01 Rev D
5-9
optional IEEE-488 programming card is enabled, the reference signal is routed to the reference
input through resistor R78. Resistor R10 and capacitors C7 and C8 are compensation
components for U3-D.
The current control circuit consists of op amp U3-C and its related components. Operation of
the current control circuit is similar to that of the voltage control circuit. The 0-5V output current
feedback signal from the A4 shunt amplifier is monitored at the inverting input of U3-C (pin 9).
The 0-5V reference signal, from either the front panel control (via analog gate U3-D), a remote
analog programming source (via analog gate U3-A), or an IEEE-488 programming card (via
resistor R77), is monitored at the non-inverting input (pin 10).
Resistor R45 and diode CR17 "OR" the output of the voltage and current control circuits to
derive the A3 PWM and A4 down programming control signal. During voltage mode operation,
the output of the current control circuit remains high and the output of op amp U3-D provides
the control signal through diode CR17. During current mode operation, the output of the voltage
control circuit is high, diode CR17 is reverse-biased, and the output of op amp U3-C supplies
the control signal through resistor R45. The control signal regulates the output of the supply by
providing negative feedback to the A3 PCB PWM error amplifier. For example, if the power
supply output voltage or current attempts to rise above the selected level:
•
the control signal from the voltage and current control circuits decreases and
•
causes the output of the PWM error amplifier to decrease, which
•
reduces the output pulse width from the PWM, and
•
reduces the voltage and current output to the selected level.
A complementary reaction occurs if the power supply output attempts to fall below the selected
output. The control signal, error amplifier, and PWM output pulse width all increase to raise the
output to the selected value.
Op amp U3-B monitors the voltage across diode CR17 to control the voltage and current mode
indicator LEDs on the front panel of the supply. When the supply is operating in voltage mode,
diode CR17 is forward-biased, the output of U3-B is high, and the voltage mode LED is
illuminated. When the supply is operating in current mode, diode CR17 is reverse-biased, the
output of U3-B is low, and the current mode LED is illuminated. In addition to controlling the
front panel mode LEDs,
op amp U3-B also provides an analog remote monitoring signal at J3
connector pin 5 and a readback signal for the optional IEEE-488 card at pin 6 of connector P2.
The internal shutdown/soft start circuit is made up of MOSFET Q1, capacitor C3, diode
CR3, and resistors R2 and R3. The circuit is activated by the remote shut down, OVP,
over-temperature shutdown, and sense line protection circuits, as well as by the front panel
standby switch. When any of these circuits are activated, a high signal is applied through diode
CR3 to charge capacitor C3 and turn ON Q1. This clamps the reference input for the voltage
control circuit at pin 12 of U3-D to ground and causes the output of the supply to fall to zero.
When the shutdown condition is removed, capacitor C3 discharges through resistors R2 and
R3 and the voltage control reference at pin 12 makes a controlled rise.
Содержание Sorensen DC Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 8: ...iv This page intentionally left blank...
Страница 10: ...vi This page intentionally left blank...
Страница 12: ...viii This page intentionally left blank...
Страница 24: ...Introduction Sorensen DCS E Series 3kW Supplies 1 6 M362295 01 Rev D This page intentionally left blank...
Страница 38: ...Installation Sorensen DCS E Series 3kW Supplies 2 14 M362295 01 Rev D This page intentionally left blank...
Страница 44: ...Basic Operation Sorensen DCS E Series 3kW Supplies 3 6 M362295 01 Rev D This page intentionally left blank...
Страница 70: ...Advanced Operation Sorensen DCS E Series 3kW Supplies 4 26 M362295 01 Rev D This page intentionally left blank...
Страница 86: ...Theory of Operation Sorensen DCS E Series 3kW Supplies 5 16 M362295 01 Rev D This page intentionally left blank...
Страница 92: ...Maintenance Sorensen DCS E Series 3kW Supplies 6 6 M362295 01 Rev D This page intentionally left blank...