Operation 

SW 5550A•SW 3700A•SW 1850A 

3-24 

Operation Manual 

SYNC OUT

 selects the output sync signal to ON, OFF or EVENT.  In the 

ON

 position, 

the signal is driven by the output frequency of the waveform, or sequence settings.  In 

the 

EVENT

 position, a pulse will be generated for any change in VOLT, CURR, FREQ, 

PHASE or WAVEFORM. 

The 

SYNC OUT

 setting only affects normal (non-sequence) mode.  To set up the SW 

so that a sync pulse only occurs for a specified sequence segment, set this parameter 

to OFF.  That way, before and after a sequence is run, no other sync pulses will occur. 

Table 3–2 provides self-test result definitions. 

BIT 

MEANING 

BIT 

MEANING 

0001 

LCD Test 

0100 

Battery Test 

0002 

EPROM CKSUM 

0200 

QSPI Test 

0004 

NVRAM BANK 0 CKSUM 

0400 

DFI Test 

0008 

NVRAM BANK 1 CKSUM 

0800 

Speaker Test 

0010 

RAM Test 

1000 

DWSB Self-Test 0 

0020 

VXI Test 

2000 

DWSB Self-Test 1 

0040 

QSCI Test 

4000 

T&MB Self-Test 0 

0080 

GPIB Test 

8000 

T&MB Self-Test 1 

Table 3–2  SELF-TEST Results Definitions 

 

Содержание SmartWave SW1850A

Страница 1: ...0267 Email Domestic Sales domorders sd ametek com International Sales intlorders sd ametek com Customer Service service ppd ametek com Web www programmablepower com SmartWave Switching Amplifier Opera...

Страница 2: ......

Страница 3: ...only Notice of Copyright SmartWave Switching Amplifier Operation Manual 2010 AMETEK Programmable Power Inc All rights reserved Exclusion for Documentation UNLESS SPECIFICALLY AGREED TO IN WRITING AME...

Страница 4: ...ii This page intentionally left blank...

Страница 5: ...nnel and equipment safety Always ensure that facility AC input power is de energized prior to connecting or disconnecting any cable In normal operation the operator does not have access to hazardous v...

Страница 6: ...nstructions of AMETEK is opened modified or disassembled in any way without AMETEK s consent or is used in combination with items articles or materials not authorized by AMETEK The Buyer may not asser...

Страница 7: ...5 1 Parameters Measured 1 5 1 5 2 Measurement Capabilities and Accuracies 1 6 1 6 Protection and Safety 1 9 1 7 Agency Requirements 1 9 1 8 Physical Specifications All Models 1 10 1 9 Environmental Da...

Страница 8: ...10 2 8 1 187 to 264 VRMS L L 3 Phase Operation 3 Wire USA 2 12 2 8 2 342 to 457 VRMS L L 3 Phase Operation 4 Wire INTL 2 12 2 8 3 Single Phase Input Connections 2 12 2 9 Output Connections to the Loa...

Страница 9: ...Table 2 3 EXT IN Connector Pinout 2 7 Table 2 4 DFI Connector Pinout 2 8 Table 2 5 IEEE 488 2 Connector Pinout 2 9 Table 2 6 Slave Connector Pinout 2 9 Table 2 7 Input Currents for 3 Phase Input Power...

Страница 10: ...e Rack Mount Version 1 1 Figure 2 1 Mounting Dimensions SmartWave Front View 2 3 Figure 2 2 Mounting Dimensions SmartWave Top and Side Views 2 4 Figure 2 3 SmartWave Rear Panel 2 5 Figure 2 4 Parallel...

Страница 11: ...high DC content for critical ATE and power line disturbance simulation testing The SmartWave can create complex waveforms with high DC content for simulating real world power irregularities including...

Страница 12: ...mputer PC using optional software A GPIB card in the computer is also required A library of 50 factory supplied waveshapes is provided in Flash Memory Non volatile memory storage is available for an a...

Страница 13: ...VRMS 3 L L 3 wire or 342 to 457 VRMS 3 L L 4 wire A chassis ground is also required for safety Input Power Factor 6 with USA rectifier 35 with INTL rectifier 99 with input PFC option Input Frequency...

Страница 14: ...ale for a 10 input line change Voltage Accuracy 0 1 of range Above 1 kHz add 0 2 kHz Add 0 1 of full scale for AC PLUS DC mode Valid for 5 to 156 VRMS and 10 to 312 VRMS at 25 C 77 F sense leads conne...

Страница 15: ...nt Programs 1000 user created segments stored in non volatile RAM up to 100 segments per sequence Segments include wave shape amplitude frequency phase angle time from 1 msec to 7 days and number of c...

Страница 16: ...ed range 1 5 2 3 Phase to Phase RMS Voltage Calculation Calculated from Phase to Neutral voltages and phases Range 0V to 700V Accuracy and Temperature Coefficient Same as for the Phase to Neutral volt...

Страница 17: ...and parallel mode use phase A for parallel mode Range 1 0 kW to 1 8 kW 3 phase mode Range 2 0 kW to 5 6 kW parallel mode and total 3 phase power Accuracy 2 5 of range DC or 40 to 500 Hz for crest fact...

Страница 18: ...s filter is used to filter the output signal before the zero comparator Resolution Frequency is displayed to 5 figures maximum the leading zeros are blanked Displayed resolution is 0 01 Hz Accuracy 0...

Страница 19: ...t with Timed Shutdown Can be set to 1 of range the timeout can be set from 100 msec to 10 sec Programmable Constant Current Can be set to 1 of range 0 5A to 16A for 156V range 0 5A to 8 0A for 312V ra...

Страница 20: ...9 Environmental Data Operating Temperature 0 C to 45 C 32 F to 113 F Note temperature measured at air inlet of SW If cabinet mounted cabinet inlet air may be rated at lower maximum temperature due to...

Страница 21: ...PLL input frequency range is 40 00 Hz to 5000 00 Hz Tracking range is 10 of programmed PLL center frequency External PLL input duty cycle is 50 10 External PLL input slew rate is 02 of input frequency...

Страница 22: ...em Firmware System firmware is stored in flash memory This makes it possible to upgrade the firmware without disassembling or returning the unit 1 11 Options Parallelable For Additional Power above 55...

Страница 23: ...put voltage and adjust the program value as necessary 1 12 2 Servo Boost Servo Boost is an enhancement to an RMS Servo used to improve its low voltage performance An RMS Servo uses an analog multiplie...

Страница 24: ...set off to prevent a tail from occurring during the transition out of the sequence mode due to the servo tracking to a new steady state condition See the System menu Section 3 3 7 1 12 3 X Load The ba...

Страница 25: ...prior to accepting the package from the carrier If extensive damage to the shipping container is evident a description of the damage should be noted on the carrier s receipt and signed by the driver...

Страница 26: ...er above the instrument may improve cooling The temperature of the intake air should not exceed 113 F 45 C At full power the unit dissipates over 2250W with the PFC option It is important that the hea...

Страница 27: ...outline and mounting dimensions of the unit and customer wiring conduit details Refer to Figure 2 3 and Table 2 2 through Table 2 6 for rear panel connector information TYPE MANUFACTURER PART NUMBER M...

Страница 28: ...Installation SW 5550A SW 3700A SW 1850A 2 4 Operation Manual Figure 2 2 Mounting Dimensions SmartWave Top and Side Views...

Страница 29: ...SW 5550A SW 3700A SW 1850A Installation Operation Manual 2 5 Figure 2 3 SmartWave Rear Panel...

Страница 30: ...B D FEMALE Gold Contacts AMP PN 747052 4 6 IEEE 488 2 25 POS FEMALE Gold Contacts AMP PN 554434 1 7 SLAVE 37 POS SUB D FEMALE ITT CANNON PN ADC37SOL2 8 AUX OUT 4 POS MINIFIT JR MOLEX PN 39 29 9045 9 O...

Страница 31: ...al Input Refer to Figure 2 3 Item 3 PIN MNEMONIC LEVEL 1 SHIELD CHASSIS 2 SHIELD CHASSIS 3 SHIELD CHASSIS 4 SHIELD CHASSIS 5 SHIELD CHASSIS 6 through 9 Not Used Not Used 10 EXT_IN_A 7 25 Vpeak DC to 5...

Страница 32: ...er the relay closes to indicate a fault has occurred The DFI Output Relay will remain latched until the front panel Power switch is cycled The DFI Input Signal is used to command the SW to open the Ou...

Страница 33: ...out 2 7 5 Slave Connector Refer to Figure 2 3 Item 7 PIN MNEMONIC PIN MNEMONIC PIN MNEMONIC 1 AGND_BUS 14 AGND_BUS 26 15A_BUS 2 AGND_BUS 15 ISNSC _BUS 27 AGND_BUS 3 AGND_BUS 16 ISNSB _BUS 28 5D_MAS 4...

Страница 34: ...A or SW 1850A via the rear panel connectors See Table 2 7 and Table 2 8 for input current values WARNING An overcurrent protection device i e circuit breaker is required in the building installation T...

Страница 35: ...RMS RECT USA 14A RMS Not Required 20A 20A RMS RECT INTL 14A RMS 25A RMS 20A 20A RMS Table 2 7 Input Currents for 3 Phase Input Power REQ D INPUT TERMINAL JUMPER CONNECTIONS CONNECT 1 PHASE INPUT POWE...

Страница 36: ...voltage will operate at this voltage CAUTION For units built before June 2004 the Neutral must not be broken by an external switch Severe damage to the unit may occur if Neutral is broken and phase vo...

Страница 37: ...ly refer to Figure 2 5 If remote sense is used Sense A is connected to Power A Sense B is connected to Power B Sense C is connected to Power C and Neutral Sense is connected to Neutral Power The three...

Страница 38: ...Installation SW 5550A SW 3700A SW 1850A 2 14 Operation Manual Figure 2 4 Parallel Connections Figure 2 5 Sense Lead Connections for 3 Phase Output...

Страница 39: ...your power applications These guidelines are equally applicable to both DC and low frequency AC up to 450 Hz power cabling The same engineering rules apply whether going into or out of an electrical d...

Страница 40: ...der peak voltages and currents which may be up to ten times the average values An underrated wire gauge adds losses which alter the inrush characteristics of the application and thus the expected perf...

Страница 41: ...determines which ohms 100 feet entry is required from Column 3 Read the corresponding wire gauge from Column 1 Column 3 value Vloss Ipeak x 0 02 x cable length Where Column 3 value Entry of the table...

Страница 42: ...any flame retardant heat resistant moisture resistant thermoplastic insulation rated to a nominal 75 C 167 F Voltage breakdown must exceed the combined effects of The rated output voltage Transient v...

Страница 43: ...Wave unit Context sensitive help is available from the front panel by pressing the Help key 3 2 Front Panel Controls The SW front panel is used for programming and data entry in local mode operation T...

Страница 44: ...to the previous menu level to abort editing a data entry field when using the numeric keypad or to switch from remote to local mode WAVEFORM MONITOR BNCs Trigger output provides a synchronization pul...

Страница 45: ...the key to change the polarity then press the Enter key to accept the value FUNCTION KEYS F1 THROUGH F3 Functions of these keys depend on the selected menu item and is defined in the Help screens OUT...

Страница 46: ...ace a box around the value area then use the Up Down Arrow keys or the Knob to select the data to be entered Press the Enter key to accept the data 3 3 1 MAIN MENU The MAIN MENU consists of the sub me...

Страница 47: ...INSTR SYSTEM PHASE A or B or C AMPL V CURL A FREQ Hz ANG Deg FUNC MAIN MENU OUTPUT OFF PROGRAM Menu Status and Information Data Menu Item Status and Information Line Range Default PHASE A enter for n...

Страница 48: ...C for positive and negative DC AC DC Coupling must also be selected from the INSTRUMENT menu The AMPL field is programmed to 0V as a precaution when changing to or from the DC or DC function 3 3 2 1 L...

Страница 49: ...ored recalled as a number from 0 to 49 The incr dec mode is not available for this menu Store Recall functions do not affect the Lock Unlock status of the Program menu fields If a field is locked it w...

Страница 50: ...AIN MENU OUTPUT OFF The Measure menu contains six display fields that can be set to any of the available measurements To set a measurement field 1 Move the field select cursor to one of the six measur...

Страница 51: ...waveshape in memory The waveshapes and waveforms are limited to spikes dropouts sags and surges of existing waveforms in the scratchpad area This is accomplished by loading the existing waveshape into...

Страница 52: ...ion in the scratchpad is erased when a new waveform is loaded LOAD EDIT VIEW SAVE DELETE FUNC Nam e WAVEFORM OUTPUT OFF Displays library of waveforms LOAD Sub Menu Status and Information Line Data Men...

Страница 53: ...and Information Line Range FREQ waveform frequency 40 5000 Hz Vrms output waveform voltage 0 156 or 0 312 V START starting phase angle 0 360 TIME surge sag duration 0 1 Freq STOP ending phase angle St...

Страница 54: ...EFORM OUTPUT OFF Selecting VIEW will display the waveform in the scratchpad as shown below 4 4 2 2 2 1 or 4 4 2 2 2 1 0 90 180 270 or WAVEFORM OUTPUT OFF Basic display after selecting VIEW The minimum...

Страница 55: ...e last character must be 2 Press the Enter key 3 Go to SAVE or SAVE RMS and press the Enter key to save LOAD EDIT VIEW SAVE DELETE NAME SAVE SAVE RMS WAVEFORM OUTPUT OFF SAVE Sub Menu Status and Infor...

Страница 56: ...M will be erased You must cycle power to the unit after this function for the changes to take effect NOTE Under normal operating conditions this function is not required Instead use Delete All User Wa...

Страница 57: ...3 3 5 1 LOAD Sub Menu The SEQUENCE menu allows you to LOAD an existing sequence from non volatile or flash memory to the sequence scratchpad NEW Erases all segments in the SEQUENCE scratchpad LOAD EDI...

Страница 58: ...See Section 3 4 Front Panel Programming Conventions ANGL phase angle degrees 0 360 The Step symbol refers to a voltage or frequency ste p in which the voltage or frequency value in the current segment...

Страница 59: ...Function keys F1 through F3 are required to create and edit sequences These keys are dependent on the field select cursor position and are defined as follows The SEG Edit Field The SEG edit field is u...

Страница 60: ...ion does not apply to GPIB control The AMPL Edit Field For information about the relationship between AMPL and FUNC refer to Section 3 4 Front Panel Programming Conventions 3 3 5 3 SAVE Sub Menu The S...

Страница 61: ...o the factory settings All factory sequences will be restored and all user sequences stored in non volatile RAM will be erased You must cycle power to the unit after this function for the changes to t...

Страница 62: ...utomatically programmed to 0 volts restored to the waveforms and values before the sequence began or remain at the waveforms and values of the last segment in the sequence The last two options will oc...

Страница 63: ...aded when switching between RUN STEP or STOP The sequence will begin executing immediately after a RUN or STEP command STEP Step through the previously loaded sequence Each segment will execute and re...

Страница 64: ...When the Parallel mode is ON only Phase A will appear in the Program menu Rear panel output power wiring must match the Parallel mode selection CAUTION Parallel mode must match the rear panel output w...

Страница 65: ...RESTORE DEFAULTS HINT MESSAGE ON SELF TEST 0000 SYNC OUT ON SYSTEM OUTPUT OFF USER Sub Menu Status and Information Data Menu Item Status and Information Line SAVE SYSTEM CONFIG store new power up sett...

Страница 66: ...t up the SW so that a sync pulse only occurs for a specified sequence segment set this parameter to OFF That way before and after a sequence is run no other sync pulses will occur Table 3 2 provides s...

Страница 67: ...indicate the internal communication status of the Digital Waveform Synthesis board and Test and Measurement board These should always indicate ON LINE unless a failure has occurred T MB will indicate...

Страница 68: ...iolated DC 40 5000 Hz Refer to Table 2 3 for external analog inputs EXT_IN_A EXT_IN_B EXT_IN_C Table 3 3 provides additional data on the External sub menu USER COMM EXTERNAL FAULTS CONFIG EXT MOD OFF...

Страница 69: ...her conditions requiring fast transient response The RMS servo is bypassed As a result load regulation drops off and current limit voltage foldback is disabled CLOCK LOCK Use this option to configure...

Страница 70: ...0 LOG4 00 00 00 LOG5 00 00 00 LOG6 00 00 00 SYSTEM OUTPUT OFF Fault History The FAULTS menu will display the last six shutdown faults in hexadecimal All three fault registers are displayed with LOG1 a...

Страница 71: ...C voltage is present in the output when the AC coupling is selected i e a waveform with DC content was selected when AC coupling was active Action SW system opens relays and shuts off ROV SW output ex...

Страница 72: ..._OV Monitors the external input in the Ex ternal Direct mode to ensure that it does not exceed 5 2 VRMS Action Opens relays displays error message and sets VPROGRAM to zero volts FLT_IN Fault Input fr...

Страница 73: ...option If you enter an incorrect code the unit will beep and the message I NCORRECT RESET CODE will be displayed the code will be reset to 0 and no changes will occur to the calibration constants T MB...

Страница 74: ...mplitudes cannot be input for positive functions Positive amplitudes cannot be input for negative functions If the amplitude of a phase is positive selecting a negative polarity function will set the...

Страница 75: ...LOAD use the Knob or Up Down Arrow keys to select LOAD 3 Press the Enter key to enter the FUNC function mode 4 Press Enter again to place a box around the waveshape selection 5 Use the Knob to cycle...

Страница 76: ...ect Vrms Using the keypad enter 1 2 0 then press Enter to set the RMS voltage to 120 volts 11 Use either the Knob or the Up Down Arrow keys to select START Using the keypad enter 0 then press Enter to...

Страница 77: ...aveform on the front panel phase A output a Return to the MAIN MENU by pressing Menu twice b Select the PROGRAM Menu then press Enter 17 In the PHASE A sub menu a Select FUNC with the Knob then press...

Страница 78: ...aveform will be displayed on Phase B output on the front panel e With the field select cursor still at AMPL press Enter to place a box around the waveshape selection then using the Knob vary the ampli...

Страница 79: ...EM MAIN MENU OUTPUT OFF 1 At the MAIN MENU select the WAVEFORM menu then press Enter LOAD EDIT VIEW SAVE DELETE WAVEFORM OUTPUT OFF 2 The WAVEFORM Menu will default to the LOAD sub menu If the cursor...

Страница 80: ...t the waveform is now in scratchpad memory 7 Press Menu to return to the WAVEFORM Menu 8 Use either the Knob or the Up Down Arrow keys to select the EDIT sub menu then either press Enter or the Right...

Страница 81: ...Enter Note that the instrument automatically calculates the TIME in this case 0 93 ms 12 Select AMPL Enter 220 6 then press Enter to set the amplitude to 220 6V 13 Press Menu to return to the WAVEFORM...

Страница 82: ...WAVEFORM SEQUENCE INSTR SYSTEM PHASE A AMPL 120 00 V CURL 5 00 A FREQ 60 00 Hz ANG 0 00 Deg FUNC SCRATCH MAIN MENU OUTPUT OFF 16 In the PHASE A sub menu a Select FUNC with the Knob then press Enter to...

Страница 83: ...NSTR SYSTEM MAIN MENU OUTPUT OFF 1 At the MAIN MENU select the WAVEFORM menu then press Enter LOAD EDIT VIEW SAVE DELETE WAVEFORM OUTPUT OFF 2 The WAVEFORM Menu will default to the LOAD sub menu If th...

Страница 84: ...icate that the waveform is now in scratchpad memory 7 Press Menu to return to the WAVEFORM Menu 8 Use either the Knob or the Up Down Arrow keys to select the EDIT sub menu then either press Enter or t...

Страница 85: ...Note that the instrument automatically calculates the STOP phase angle value in this case 66 6 12 Select AMPL enter 10 for 10 volts then press Enter 13 Press Menu to return to the WAVEFORM Menu LOAD E...

Страница 86: ...A FREQ 60 00 Hz ANG 0 00 Deg FUNC SCRATCH MAIN MENU OUTPUT OFF 16 In the PHASE A sub menu a Select FUNC with the Knob then press Enter to place a box around the waveshape selection b Rotate the Knob...

Страница 87: ...programmed output frequency true rms 0 7071 40 5000 Hz DcRip03 DC with 3 ripple Frequency of the ripple is deter mined by the programmed output frequency true rms 0 7071 40 5000 Hz DcRip10 DC with 10...

Страница 88: ...0 40 5000 Hz Inrush Sine wave with first quarter 0 to 90 at 0Volts 1 0 40 5000 Hz Noisz010 Sine with 10 noise at zero crossings true rms 1 0 40 5000 Hz Noisz100 Sine with 100 noise at zero crossings t...

Страница 89: ...Six voltage step sine wave 1 0 40 5000 Hz Step_12 Twelve voltage step sine wave 1 0 40 5000 Hz Step_24 Twenty four voltage step sine wave 1 0 40 5000 Hz Taylor 5th harmonic Taylor series wave true rms...

Страница 90: ...Standard Waveforms SW 5550A SW 3700A SW 1850A 4 4 Operation Manual This page intentionally left blank...

Отзывы: