5.1.4
Ball & Nozzle Cleanliness Check
The ball should pass smoothly through the nozzle bore without any restriction or friction. If
resistance is felt, reclean the ceramic ball and nozzle and repeat cleanliness check.
5.1.3
Cleaning Ball and Nozzle
AMETEK recommends cleaning the ceramic ball and nozzle with a residue-free solvent.
Using a cottom swab or soft cloth with a solvent such as alcohol,
wipe the inside of the nozzle.
Dampen a clean cloth with alcohol or other residue-resistant
solvent, and clean the ceramic ball by rotating it within the cloth.
After the ball and nozzle have been cleaned, slide the ball through
the nozzle. The ball should pass through the nozzle with no
resistance.