AMETEK CTS
PCD 8 s-series
Operating Manual
V 2.02
14 / 53
Assembly for ANSI coupling or additional IEC lines coupling mode
For ANSI coupling or additional IEC lines coupling add 2 coupling modules SCB on the big positioning plate and
press them slightly down into the positioning holes. Fix the SCB
’s by pushing the black quick snap-on knobs
down (A). Connect the ground terminals of both SCB
’s to the decoupling networks (B).
NOTE Because of the coupling capacitors, the mains voltage can be present at the HV connectors
when the EUT power supply is switched on. The HV connectors must therefore always be
hooked up to the generator before the mains and the EUT power supply are switched on.
NOTE The whole test assembly should be supplied from the same mains connection in order to
prevent uncontrolled flow of pulse current in other parts of the system. Installation in a
Faraday cage ensures that non-associated items and equipment are not disrupted by pulses
radiated from the cabling or the device under test. Connections to the EUT must be of low
impedance and be made with high contact pressure. Otherwise welding or arcing might occur
at the contact points.
3.2.4. Grounding
The instruments conform to the safety requirements, but with an increased leakage current given by the
decoupling filter. Operate the equipment only in dry and clean surroundings. Any condensation that occurs must
be allowed to evaporate before putting the equipment into operation. Do not exceed the permissible ambient
temperature or humidity given in the IEC specification.
NOTE
Operation without a protective earth connection is forbidden!
NOTE
Only handle the test rig or the device being tested when the EUT power supply is switched off and the
generators test sequence has been halted or aborted.
Содержание PCD 8 S Series
Страница 37: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 37 53 ...
Страница 38: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 38 53 9 Coupling guide 9 1 IEC coupling Line to Ground 9 µF ...
Страница 39: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 39 53 9 2 IEC coupling Line to Line 18 µF ...
Страница 40: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 40 53 9 3 Optional IEC coupling Lines to Ground 9 µF 1 3 ...
Страница 41: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 41 53 9 4 Optional IEC coupling Lines to Ground 9 µF 2 3 ...
Страница 42: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 42 53 9 5 Optional IEC coupling Lines to Ground 9 µF 3 3 ...
Страница 43: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 43 53 9 6 Optional ANSI coupling 1 5 Basic ...
Страница 44: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 44 53 9 7 Optional ANSI coupling 2 5 Basic ...
Страница 45: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 45 53 9 8 Optional ANSI coupling 3 5 Supplemental ...
Страница 46: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 46 53 9 9 Optional ANSI coupling 4 5 Supplemental ...
Страница 47: ...AMETEK CTS PCD 8 s series Operating Manual V 2 02 47 53 9 10 Optional ANSI coupling 5 5 Diagnostic ...
Страница 52: ......
Страница 53: ...AMETEK CTS PCD 8 s series Manual of operation V 1 00 53 53 ...