![Ametek Mocon Dansensor LeakPointer 3 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/ametek/mocon-dansensor-leakpointer-3/mocon-dansensor-leakpointer-3_user-manual_2931374037.webp)
Dansensor
®
LeakPointer 3/3
+
EN
User Guide
35
6. In the appearing window...
...enter
100
as the current test hole size and press
key.
7. As you can see in the appearing
Leak Correction Factor
window, it takes 5
measurements to calculate a valid correction factor.
8. Close the chamber to start the measurement.
9. When the test is completed, the result is displayed in the corresponding field, an
Average
value is calculated and the value in the
Leak gain factor
field is updated accordingly.
10. Prepare a new package as described earlier, place it in the chamber and start the next
measurement.
Содержание Mocon Dansensor LeakPointer 3
Страница 1: ...P N 380064 F 06 2021 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Страница 8: ...6 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Страница 20: ...18 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Страница 94: ...92 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Страница 98: ...96 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Страница 99: ...This blank page has been inserted to enable double sided printing of the document...
Страница 100: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...