103887 01 

 

7

 

Press 

 

for automatic changing of  current value. If STEP is selected 

            (the AUTO-STEP function), change-over takes place automatically between the pre-

defined current values. For proportional transmitters (VALVE), change-over takes 
place between 3.8 - 4 - 4.2 - 4..... and 19 - 20 - 21 - 20..... mA 

 

(the range is selected using the “up arrow” or “down arrow” keys). 

 
The mAcal now generates the current shown in the display. 

 

3.4    HOW THE CURRENT IS MEASURED 

Connect the mAcal to the transmitter to be tested as shown in Fig. 5 or Fig. 6 on the cover 
of the user’s guide and switch the unit on. 

 

Press 

 

until the display shows 

MEASURE mA

 or 

MEASURE %

 

 

 

 

Press 

 

for selection of INT or EXT (the  present selection is 
shown in the display). 

 

 

 
If INT is selected, the +24VDC of the current loop is generated by the mAcal itself. If EXT is 
selected, the +24VDC of the current loop is generated by an external power supply. 
 

 

Press 

 

for selection of readout in mA or %. 

 
The display thus has the following appearance in the measurement of 20mA in the current 
loop. 
The current is printed out with a resolution of 0.001mA: 

 

 
The mAcal now measures the current shown in the display. 
 

3.5    HOW THE VOLTAGE IS MEASURED 

Connect the mAcal to the instrument to be tested as shown in Fig. 7 on the cover of the 
user’s guide and switch on the unit. 
 

 

Press 

 

until the display shows 

MEASURE V

.  Readout in % 

cannot be selected. 

   

 

 
The display thus has the following appearance in the measurement in V. The voltage is 
printed out with a resolution of 0.1Volt:  

Содержание mAcal

Страница 1: ...User Manual mAcal mA loop Calibrator Copyright 2005 AMETEK DENMARK A S...

Страница 2: ...SURED 7 3 6 HOW NEW PARAMETERS ARE SELECTED SETUP 8 3 6 1 SELECTION OF MODE 8 3 6 2 SELECTION OF SLOPE TIME 8 3 6 3 SELECTION OF STEP SIZE 9 3 6 4 SELECTION OF STEP TIME 9 3 6 5 SELECTION OR DE SELECT...

Страница 3: ...ment In order to avoid any personal injuries and or damage to the instrument all safety instructions and warnings must be observed Disposal WEEE Directive These calibrators contain Electrical and Elec...

Страница 4: ...e contacts Keep your fingers behind the finger guards on the probes Do not use the calibrator if it operates abnormally Protection may be impaired When in doubt have the calibrator serviced Disconnect...

Страница 5: ...mA MEASURE mA and MEASURE Volt function PWR POWER Change over takes place between EXT the 24VDC of the current loop is generated by an external power supply and INT the 24VDC of the current loop is ge...

Страница 6: ...s the readout in as well as STEP values 4 20 4 20mA is selected as the range which affects the readout in as well as STEP values FLOW FLOW transmitters are selected which affects the readout in as wel...

Страница 7: ...tested as shown in Fig 3 or Fig 4 on the cover of the user s guide and switch the unit on Press until the display shows SOURCE mA or SOURCE Press for selection of INT or EXT the present selection is...

Страница 8: ...n of INT or EXT the present selection is shown in the display If INT is selected the 24VDC of the current loop is generated by the mAcal itself If EXT is selected the 24VDC of the current loop is gene...

Страница 9: ...nearly FLOW calculated logarithmically VALVE not calculated AUTO STEP function 3 8 4 4 2 4 or 19 20 21 20 mA Press or for selection of new MODE Press to save the present selection 3 6 2 Selection of S...

Страница 10: ...between changes in the AUTO STEP function can be changed Either s or s is shown in the display Press or for setting of STEP TIME Press to save the present selection 3 6 5 Selection or de selection of...

Страница 11: ...he text flashes in the display CAUSE The unit measures a negative current REMEDY The test leads and are probably to be changed over 4 3 TEST FOR OVERRANGE The text flashes in the display CAUSE The uni...

Страница 12: ...instrument can still be used LOW BAT flashes The batteries are used up The instrument cannot be used The display goes out However the keyed in values are retained REMEDY LOW BAT lights up Change the b...

Страница 13: ...e shows the software version After approx 1 second the serial number of the pcb is shown After another approx 1 second the text is shown code must be keyed in Press to exit service mode This option ca...

Страница 14: ...til the display shows the same value as is measured with the external precision instrument Press to approve the value The mAcal now carries out a self calibration and adapts its circuits to that measu...

Страница 15: ...mA output internal loop pwr Battery types 4 types LR6 AA 1 5V Alkaline Mains adapter 103950 6V DC 230mA EURO with 1 3 mm DC female connector 103957 6V DC 230mA UK with 1 3mm DC female connector 103964...

Отзывы: