background image

User Guide

Landcal Blackbody Source

Type R1200P

Page 9

6  Using the Furnace

6.1  Introduction

The furnace has been designed to create a target of uniform temperature with 

a high emissivity at short wavelength for calibrating mainly silicon cell and 

germanium detector radiation thermometers.  The target is a conical cavity at 

one end of a hollow cylinder, and the thermometer views the target along the 

axis of the cylinder.
The cone plays an important part in establishing uniformity of radiation emitted 

by the target, and when calibrating radiation thermometers the target size 

requirements must be met by the cone.  If the thermometer target size is so big that 

the walls of the cylindrical block are viewed, a less accurate calibration will result.

6.2  Measuring thermocouple/indicator

A thermocouple embedded in the conical target is used to measure the 

temperature.  The “TARGET” digital indicator on the side of the furnace displays 

the temperature measured.
Agreement between the thermocouple measurement and a reference standard 

radiation thermometer should be better than ±10°C.
The errors are repeatable, and a correction curve can be obtained with a 

Certificate of Calibration for the source.  A UKAS (United Kingdom Accreditation 

Service  (National Measurement Accreditation Service) Certificate of Calibration 

is available from LAND Instruments International.

6.3  Control thermocouple/controller

A second thermocouple is used to control the furnace temperature.  As part of 

the control strategy, this thermocouple is situated close to the furnace heating 

elements.  As a result the control thermocouple is actually hotter than the 

measurement thermocouple.
The setpoint on the furnace controller will invariably need to be set higher than 

the desired target temperature.  The difference between setpoint and target 

temperature can be evaluated for a given furnace by experiment, always using 

the TARGET indicator as a measure of target temperature.

6.4  On-site use

The source is portable, so can be used either in the laboratory where the 

radiation thermometers are brought to the source for checking, or taken on-site 

where the source is used close to the thermometer installations.
Due to the lightweight construction of the source damage may occur if it is 

mishandled.   The source should always be treated with respect and handled 

in the same way as any other scientific instrumentation.   Following the 

precautions below will extend the life of the source.

Содержание Landcal R1200P

Страница 1: ...le 01246 410585 Email land infrared ametek co uk Internet www landinst com Land Instruments International 2008 2012 AMETEK Land Inc 150 Freeport Road Pittsburgh PA 15238 U S A Telephone 412 826 4444 F...

Страница 2: ...at safety precautions are taken whenever the materials are handled Exposure to fibrous dust may cause respiratory disease When handling the material always use an approved respiratory protection equip...

Страница 3: ...n It is possible that this equipment is to be installed on or near to machinery or equipment operating at high temperatures and high pressures Suitable protective clothing along with face and eye prot...

Страница 4: ...s etc Design and Manufacturing Standards The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design manufacture and on site servicing of combustion en...

Страница 5: ...Source Type R1200P 1 2 Description 3 3 Specification 4 4 Electrical Supply details 6 5 Commissioning 7 6 Using the Furnace 9 7 Calibration of Radiation Thermometers 11 8 Maintenance 12 9 Eurotherm Te...

Страница 6: ...Landcal Blackbody Source Type R1200P User Guide Page 6...

Страница 7: ...the thermometer can be conducted by using the target indication From previous work the target indication temperature has been found to agree with the radiance temperature to within 10 C at short wave...

Страница 8: ...some individuals be an irritant to the skin nose and throat Safety Note Refractory Fibrous Insulation This product contains alumino silicate wool products in its thermal insulation These materials ma...

Страница 9: ...Fig 1 Landcal Blackbody Source Type R1200P CA970259 If the source is used as a transfer source i e calibration by comparison method adopted a set of optional standard radiation thermometers will be re...

Страница 10: ...roximately 0 98 at short wavelengths Themocouple type Both the Control thermocouple and Indicator thermocouple are Type N NiSil NiCrNiSil mineral insulated 3mm OD with Nicrobell B sheath 350mm long pl...

Страница 11: ...y R Red Element TC N L Indicator Controller Voltage change over switch shown in 220 240 position Panel Mounted Fuse 10A Super Quick Blow FF BU BU BU BU R R R GR Y Green Yellow R R PI G GR Y 4 6 2 3 5...

Страница 12: ...nnected from the supply The isolator or the plug must be within easy reach of the operator The supply MUST incorporate an earth ground The wiring diagram for the R1200P is given in Fig 2 The power cab...

Страница 13: ...the source Do not obstruct any of the vents in the control section they are needed to keep the controls cool When the source is operating at any temperatures above ambient the front plate and case wil...

Страница 14: ...n or 25 pin to connect to the computer RS232 Cable Furnace to PC connections Furnace end of cable female Computer end of cable male 9 pin 9 pin 25 pin Rx 3 3 2 Tx Tx 2 2 3 Rx Com5 5 7 Com 7 8 4 5 Link...

Страница 15: ...curve can be obtained with a Certificate of Calibration for the source A UKAS United Kingdom Accreditation Service National Measurement Accreditation Service Certificate of Calibration is available fr...

Страница 16: ...e and carry the source in its accessory carrying case 3 After use it is recommended that the source be left to cool to an internal temperature of 200 C before moving 4 Never place the source into the...

Страница 17: ...ing temperature of the thermometer to be tested The target temperature is that indicated by the measuring thermocouple displayed on the target indicator 7 2 Thermometer calibration When the source is...

Страница 18: ...f the heating element and insulation 8 2 Fitting a replacement heating element insulation package The location of the terminal blocks on the insulation assembly is shown in Fig 3a The replacement asse...

Страница 19: ...which terminal 7 Remove the insulation assembly from the base 4 set screws on the underside complete with earth connection cable To fit a new insulation assembly 1 Attach the insulation assembly onto...

Страница 20: ...ff white paste to give good heat transfer to the aluminium sheet New relays are supplied with a heatsink pad All remnants of the old heatsink paste must be removed The thin white heatsink pad must be...

Страница 21: ...trols when the set point is reached 8 5 Storage and carrying case A custom built aluminium storage and transportation case is available as an optional extra Use of this case is recommended when a The...

Страница 22: ...res contain factory set parameters which should not be changed The Up and Down keys are used to alter the setpoint temperature in Level 1 operation and parameter values when in Level 2 operation To en...

Страница 23: ...ramp rate allowing heating and cooling at the maximum rate OFF is the default setting 9 4 Altering the Power Limit when applicable 1 Press the Scroll key until OP Hi Output High is displayed 2 Press...

Страница 24: ...Spare control or measuring thermocouple 135 172 11 Accessories The following calibration accessories are available for use with the Landcal type R1200P Description Land Part Number Storage carrying c...

Страница 25: ...ters accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VDT U DTT FLT5 A This 36 month warranty i...

Страница 26: ...onditions outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral p...

Отзывы: