IntelliPoint RF RML Series

 

Control Drawings

36

7.1  FM Control Drawings (Continued)

Содержание IntelliPoint RF RGL Series

Страница 1: ...Installation and Operating Instructions IntelliPoint RFTM RML Series Point Level Switch with Manual Calibration Set Point U S and Canada 1 800 553 9092 24 Hour Service 1 800 527 6297 International 1...

Страница 2: ...ained in this manual including but not limited to implied warranties or fitness for a particular purpose Drexelbrook shall not be liable for errors contained herein or for incidental or consequential...

Страница 3: ...int RMLXX1 LM EDO 11 08 107 Issue 11 205 Keith Valley Road Horsham PA 19044 U S and Canada 1 800 553 9092 24 Hour Service 1 800 527 6297 International 1 215 674 1234 Fax 1 215 674 2731 E mail drexelbr...

Страница 4: ......

Страница 5: ...Extend your warranty from 12 months to 36 months by registering your purchase Please go to www drexelbrook com Click on Contact Us then Warranty Registration and fill out the form...

Страница 6: ...Contents...

Страница 7: ...3 3 Calibration Procedures 15 Section 4 Spare Parts List 21 4 1 Spare Parts List 21 Section 5 Troubleshooting 23 5 1 Testing Sensing Element 23 5 2 Testing Electronic Unit 24 5 3 Testing Relay Circuit...

Страница 8: ...Section 1...

Страница 9: ...RF Admittance level transmitter is the next generation Although similar to the capacitance concept IntelliPoint employs a radio frequency signal and adds the Cote Shield circuitry within the Electron...

Страница 10: ...0202 036 int rem 3 4 bar 149 C 50 PSI 300 F 316SS and TFE 18 Corrosive material 700 0001 022 int rem 69 bar 38 C 1000 PSI 100 F TFE higher pressure with 34 5 bar 149 C 500 PSI 300 F waterlike viscosi...

Страница 11: ...6SS and Ceramic 68 9 bar 260 C 1000 PSI 500 F 61 High Temperature 700 0204 002 remote 0 bar 816 C 0 PSI 1500 F 316SS and Ceramic 62 High Pressure Temp 700 0204 048 remote 275 8 bar 316 C 4000 PSI 600...

Страница 12: ...Loop 2 External Lid Lock 2 M4 2 5 mm Hex Drive Conduit Entries 2 inch NPT FM CSA Systems M20 x 1 5 CENELEC Systems LABEL see below Equipotential Connection Figure 1 3 Dual Compartment Housing Detail T...

Страница 13: ...be between 30 to 70 C 22 to 158 F The IntelliPoint RF utilizes a dual compartment housing and a completely encapsulated input output module to reduce the possibility of damage occuring from water mig...

Страница 14: ...at electronics must not exceed 70 C 158 F When installing flange mounted sensing elements keep mating surfaces and bolts free of paint and corrosion to ensure proper electrical contact with vessel Av...

Страница 15: ...pply to all flush on off sensing elements models 700 0207 001 700 0207 002 700 0207 003 700 0207 004 700 0207 006 These systems will sense presence of material no flow or plugged chute and absence of...

Страница 16: ...ectrical power at the source Ensure the wiring electrical fittings and conduit connections conform to electrical codes for the specif ic location and hazard level The IntelliPoint RF instrument utiliz...

Страница 17: ...is at zero seconds when in the full counter clockwise position Do not force the pot past the stop or damage will occur LLFS FAILSAFE HLFS LLFS FAILSAFE HLFS TIME DELAY 0 60 SECONDS 0 60 FWD TD REV FWD...

Страница 18: ...ollow RELAY 1 providing a DPDT alarm relay Fault RELAY 2 is used to indicate a fault Instrument is supplied with RELAY 2 jumper set in alarm position RELAY 2 assignment is field selectable using a jum...

Страница 19: ...sensing element circuit board side of the housing One is used to indicate that the unit has power The remaining two LEDs are used to indicate the condition of relay 1 and relay 2 See Figure 2 6 Figure...

Страница 20: ...ing element is sealed to the housing and cannot be removed without permanent damage For IntelliPoint RF instruments that are mounted remotely from the sensing element the cable connections from the se...

Страница 21: ...ing Element Connection continued For IntelliPoint RF instruments that are mounted remotely from the sensing element the cable connections from the sensing element to the electronic unit are made to th...

Страница 22: ...Section 3...

Страница 23: ...n chatter on those occasions when level happens to be right at switching point or when surface is agitated RML Series level control with adjustable differential allows user to determine amount of capa...

Страница 24: ...uick Calibration Chart Material Being Measured Number of Clockwise Adjustment Turns from Operate Position RML Series Standard Sensitivity RGL Series High Sensitivity Conductive Materials Water Based S...

Страница 25: ...the number of turns that were counted in step E G Record that half number of turns as PRELOAD for use later in recalibration See Section 3 3 4 Calibration is Complete If less than 1 4 turn of adjustme...

Страница 26: ...internal microprocessor records and stores the Air value This is the uncovered value of the sensor mounted in the vessel The IntelliPoint will also store the last covered value and the last uncovered...

Страница 27: ...ranulars containing moisture 1 4PF One half 1 2 turn for dry insulating powders Calibration is Complete 1 4 TURN MAX ON TO OFF DRY INSULATING POWDERS MOIST GRANULARS RED LED OFF RED LED ON 3 3 5 Calib...

Страница 28: ...t operating correctly Please consult factory service department for assistance This is a simple function test of the electronics D If above operation is satisfactory then continue turning adjustor clo...

Страница 29: ...1 Section 4 Spare Parts List 4 1 Spare Parts List O ring 250 1 75 Housing inch NPT Conduit Entry 260 2 540 Housing M20 Conduit Entry 260 2 542 Input Output Module 385 48 6 Circuit Board Contact Factor...

Страница 30: ...Section 5...

Страница 31: ...sured Resistance Sensor dry and clean Center Wire Shield Center Wire Ground Ohms Shield Ground Resistance readings must be taken using an analog ohmeter set to Rx1000 scale Ohms Ohms When tank level i...

Страница 32: ...state 5 If IntelliPoint changes state while touching test point but not when touching tip of sensor in most cases integral cable is faulty Refer to Section 5 6 Testing Integral Cable 6 If IntelliPoint...

Страница 33: ...instrument has detected that pad capacitor value is toolarge Consult factory for pad capacitor values 5 6 Testing Integral Cable M3 2 5mm Hex Drive 8 7 6 5 4 2 1 3 M3 2 5mm Hex Drive M5 Slotted Drive...

Страница 34: ...EAD 0 OHMS CHECK FOR CONTINUITY 8 0 0 0 0 OHMMETER Common data NC NC 5 Power Supply Voltages 5 5 8 Testing Power Supply Power supply can be tested separately as follows A Remove power from electronic...

Страница 35: ...der Number Material being measured Temperature Pressure Agitation Brief description of the problem Checkout procedures that have failed 5 10 Field Service Trained field servicemen are available on a t...

Страница 36: ...lose all chemicals and decontaminate the sensing element To obtain a Return Authorization Number RA contact the Service department at 1 800 527 6297 US and Canada or 1 215 674 1234 Please provide the...

Страница 37: ...aximum Relay Contacts 2 SPDT Maximum Contact Load 5A 30 VDC 5A 250 VAC Maximum Switching Capacity 2000 VA 150 Watt Minimum Contact Load DC 100 mA 12 VDC 0 to 200 mA 12 VDC Optional Housing Electronics...

Страница 38: ...Section 7...

Страница 39: ...31 IntelliPoint RF RML Series Control Drawings Section 7 Control Drawings 7 1 FM Control Drawings...

Страница 40: ...IntelliPoint RF RML Series Control Drawings 32 7 1 FM Control Drawings Continued...

Страница 41: ...33 IntelliPoint RF RML Series Control Drawings 7 1 FM Control Drawings Continued...

Страница 42: ...IntelliPoint RF RML Series Control Drawings 34 7 1 FM Control Drawings Continued...

Страница 43: ...35 IntelliPoint RF RML Series Control Drawings 7 1 FM Control Drawings Continued...

Страница 44: ...IntelliPoint RF RML Series Control Drawings 36 7 1 FM Control Drawings Continued...

Страница 45: ...37 IntelliPoint RF RML Series Control Drawings 7 1 FM Control Drawings Continued...

Страница 46: ...IntelliPoint RF RML Series Control Drawings 38 7 1 FM Control Drawings Continued...

Страница 47: ...39 IntelliPoint RF RML Series Control Drawings 7 1 FM Control Drawings Continued...

Страница 48: ...IntelliPoint RF RML Series Control Drawings 40 7 1 FM Control Drawings Continued...

Страница 49: ...41 IntelliPoint RF RML Series Control Drawings 7 1 FM Control Drawings Continued...

Страница 50: ...IntelliPoint RF RML Series Control Drawings 42 7 2 Mounting and Wiring for Spark Protector Drawings...

Страница 51: ...43 IntelliPoint RF RML Series Control Drawings 7 2 Mounting and Wiring for Spark Protector Continued...

Страница 52: ...IntelliPoint RF RML Series Control Drawings 44 7 2 Mounting and Wiring for Spark Protector Continued...

Страница 53: ...45 IntelliPoint RF RML Series Control Drawings 7 3 Adding a Padded Capacitor...

Страница 54: ...IntelliPoint RF RML Series Control Drawings 46 7 3 Adding a Padded Capacitor Continued...

Страница 55: ...47 IntelliPoint RF RML Series Control Drawings 7 3 Adding a Padded Capacitor Continued...

Страница 56: ...Appendix A...

Страница 57: ...atic discharge 5 Protect electronics from any heat Shortening The bare metal center rod of the sensing element can be shortened with a hacksaw Be careful not to cut either of the two insulators See Fi...

Страница 58: ...not transfer any intellectual property rights to Buyer unless Seller specifically agrees to do so in writing Seller shall retain ownership of all applicable patents trademarks copyrights and other int...

Страница 59: ......

Страница 60: ...d Horsham PA 19044 U S and Canada 1 800 553 9092 24 Hour Service 1 800 527 6297 International 1 215 674 1234 Fax 1 215 674 2731 E mail drexelbrook info ametek com Website www drexelbrook com An ISO 90...

Отзывы: