User Manual
4
SELECTION SWITCHES ON DIP SWITCH (SEE TABLE 1)
• Switch the Power Supply switch (Pin no. 1 on DIP Switch), to the ON position.
• Select the desired Temperature Sensor by using Pin no. 2 of the DIP Switch as per the table given below.
• If Thermocouple Sensor is selected on Pin no. 2, then the type of Thermocouple may be selected by Pin no. 3 of
the DIP Switch.
• Select the desired temperature units between ° F and ° C by means of Pin no. 4 of the DIP switch.
• The display will update to the selected units after the measurement cycle of approximately ten seconds.
• The indicator will immediately indicate the process temperature on the LCD display.
BATTERY REPLACEMENT
• Two nos. Batteries (type AA, 1.5V).
• The instrument has minimal power requirements. However, when the LOW BATT sign appears on the display, it is
time to replace the batteries.
• When LOW BATT is indicated, or if the instrument has been unused or Power OFF condition for an extended
period, the batteries must be replaced immediately.
USER SETTINGS
• On the rear card, there is a four-way DIP switch. The settings are as follows:-
Table 1 - Selection Switches
Pin Number
Function in “Off” Position
Function in “On” Position
1
Power Off
Power On
2
RTD Sensor
Thermocouple Sensor
3
Thermocouple J type
Thermocouple K type
4
Degrees Fahrenheit
Degrees Centigrade
Model DT-8300EX
NOTES:
• The Model DT-8300EX has a scan Cycle Time of 10 seconds to extend Battery life. This means that the LCD
display is updated every ten seconds. Because of this, the keys may appear to respond slowly at times.
BATTERY OPERATED DIGITAL TEMPERATURE INDICATOR
17348
____________________________________________________________________________________________
03
CONFIGURATION
All settings and adjustments must be done from the rear side of the unit.
CONNECTIONS
•
Locate the green terminals on the lower rear section of the unit.
•
Connect the RTD (3 wire Pt 100) or the Thermocouple sensor input at these terminals in the correct polarity.
•
Minimize the number of junctions between the Thermocouple and the Indicator.
PROGRAMME USING ONLY 4 DIP SWITCHES
•
The DIP Switch settings are as follows :-
PIN
NO.
FUNCTION
IN “OFF” POSITION
FUNCTION
IN “ON” POSITION
1
Power Off
Power On
2
RTD Sensor
Thermocouple Sensor
3
Thermocouple J type
Thermocouple K type
4
Degrees Fahrenheit
Degrees Centigrade
•
Switch No. 1 is used to turn the unit ON or OFF. This switch is provided to save battery life when the unit is not in use for an
extended period of time.
•
Switch No. 2 is a selector switch which allows the user to select between a RTD sensor (Pt 100) input or a Thermocouple input. If
an RTD is used, this switch should be in the “OFF” position.
•
Switch No. 3 allows selection between a Type J or Type K thermocouple input, when switch no. 2 is in the “ON” position. This
switch is not used when RTD sensor (Pt 100) is selected.
•
Switch No. 4 is a selector switch which allows the user to display temperature in either degrees C or degrees F. The “OFF”
position of this switch will display degrees F.
ON
▌ ▌ ▌ ▌
1 2 3 4