CALIBRATION

Factory Adjustment

If adjustment is required, we recommend returning the unit to the factory. Factory service offers benefits you won’t find anywhere else. We have the facilities 
to test your gauge at a variety of temperatures utilizing NIST traceable standards in our A2LA accredited laboratory, resulting in calibration certificates that 
provide performance data over temperature. Furthermore, upgrades may be available to add or enhance operating features. We designed the product to last, 
and we support it so that you can get the most from your investment.

Calibration Frequency

Under normal operating conditions, we recommend the XP2i be calibrated on an annual basis. Your quality system may require more or less frequent calibra-
tion, or your experience with the gauge, or operating environment may suggest longer or shorter intervals. 

Although we prefer that you return the XP2i to Crystal Engineering for calibration, ordinary recertification and/or adjustments may be performed by any 
qualified personnel with appropriate training and equipment. 

Span Factor

There are no internal potentiometers. The XP2i contains a “span” factor, set to approximately 1 (as shipped from the factory). As components age this may 
need to be changed to a value slightly higher or lower, to slightly increase or decrease all readings. This adjustment can be made with or without a computer 
(see 

ConfigXP Configuration Software

). 

 

X

Span Factor adjustment

Note:

  The following instructions are ONLY intended for such qualified personnel with appropriate test equipment. We recommend that the calibration stan-

dards used have a minimum rated accuracy of 0.025% of reading, or equivalent in terms of percent of full scale. This level of accuracy requires the use 
of piston (deadweight) gauges or very high performance pressure controllers.

“Zero” the XP2i, then record displayed pressure for two or more pressure points. Determine if the XP2i would benefit from an overall increase or decrease of 
the indicated pressures. 

To change the span factor from the keypad, turn off the XP2i, then press the 

(

on/off

)

(

units

)

, and 

(

peak

)

 buttons simultaneously. The firmware version will 

be briefly displayed, followed by the word 

cal

, followed by the actual span value. The span factor may be adjusted by pressing either the 

(

units

)

 or 

(

peak

)

 

button to increase or decrease the value, respectively. The value changes in 0.0001 increments. Press the 

(

zero

button to store the new value in memory, or 

the 

(

on/off

button to cancel the change.

For absolute XP2is, it is possible to correct for long term drift using a second calibration factor, zero value offset. ConfigXP and a barometric reference with ac-
curacy of 0.1 PSI or better is required to perform the calibration. To calibrate the zero offset, clear the zero as described earlier in this manual by pressing and 
holding the 

(

zero

button until

 

 ) appears. Once cleared, subtract the displayed pressure from barometric pressure, add this difference to any existing 

zero value offset in ConfigXP, and update the gauge (new value = barometric – dis existing). For example, if the displayed value is 14.5 PSI, barometric 
pressure is 14.7 PSI, and the existing zero value offset in ConfigXP is 0.1 PSI, the new zero value offset would be 0.3 PSI (14.7 - 14.5 + 0.1 = 0.3).

 The span factor and zero value offset can be viewed and set directly by ConfigXP. Span factor adjustment through the keypad 

can be disabled by ConfigXP through the disable span factor feature or by password protecting the XP2i

 

.

XP

Содержание Crystal Pressure WT Series

Страница 1: ...WT Series Operation Manual for the WT Series Digital Test Gauge...

Страница 2: ...losure 9 Serial Numbers 9 Specifications 10 Power 11 Pressure Ranges Display Scales Resolution 12 Part Numbering System 12 Support 13 Troubleshooting 13 Calibration 14 Software 15 Replacement Parts 15...

Страница 3: ...hours of continuous operation The batteries are easily replaced by removing the two screws Other features include Continuous capture of peak and valley pressure readings PSV mode Programming interface...

Страница 4: ...ent That is they indicate the dif ference between applied pressure and ambient barometric pres sure However the zero button can be used to force a WT Series to read zero pressure at any applied pressu...

Страница 5: ...the peak button causes the display to cycle through the following depending on your setting in ConfigXP No Icon Live Pressure display Maximum detected pressure Minimum detected pressure blinking PSV...

Страница 6: ...ill be ignored and HI will be displayed X X To Zero the WT Series Turn off peak indication by pressing the peak button repeatedly until the Peak High icon HI and Peak Low icon LO are off Then press th...

Страница 7: ...button again the backlight will go into a lower brightness setting to extend battery life and remain on for 2 minutes after the last key is pressed Press the backlight button once more and the light...

Страница 8: ...t and set the desired density of water Note If the WT Series is password protected or inches of water is disabled you will not be able to view or change the water density from the keypad Overpressure...

Страница 9: ...ng the closing pressure of a PSV Press the peak button once to view the captured reseat pressure The Peak Low icon LO will flash on the display To clear the peaks press the clear button while the disp...

Страница 10: ...surements will be possible until the batteries are replaced Battery Replacement The WTSeries gauge uses 3 AAA alkaline batteries The batteries are located above the gauge behind a metal panel With a s...

Страница 11: ...PSI BATTERY ACCESS 3 X AAA 10 66 9 25 BOLT CIRCLE 8 70 BOLT CIRCLE 10 125 SERIAL PORT WITH COVER PRESSURE PORT POSITION 120 3 PL 3 80 0 28 THRU ALL 3 PL 8 00 REAR COVER DIAMETER IDENTIFICATION LABEL...

Страница 12: ...f temperature shock and or vibration may warrant a more frequent recertification period 20 to 100 of Full Scale 0 1 of Reading 0 to 20 of Full Scale 0 02 of Full Scale Vacuum for 300 PSI 20 bar 2000 k...

Страница 13: ...sealed Pressure Connection 1 8 female NPT Display Screen 5 5 digits Display Rate 4 readings second standard 8 readings second PSV mode Numerical Display Height 16 9 mm 0 67 single line display LCD re...

Страница 14: ...ed via the RS 232 connector using ConfigXP software kPa and MPa models can display pressure in kPa MPa and bar or mbar only PSI and bar models can display all available units WT Series will indicate p...

Страница 15: ...ason As in the previous note gas has dissolved into the hydraulic fluid When decreasing the pressure the dissolved gas then leaves the fluid but at an uneven rate so small pressure differen tial due t...

Страница 16: ...ted accuracy of 0 025 of reading or equivalent in terms of percent of full scale This level of accuracy requires the use of piston deadweight gauges or very high performance pressure controllers Zero...

Страница 17: ...adjust calibration and more via the RS 232 interface REPLACEMENT PARTS The only user replaceable parts are the batteries ACCESSORIES P N 2984 AC adapter kit Permits operation of an XP2i from an AC sup...

Страница 18: ...misused altered or damaged by accident or abnormal conditions of operation Crystal Engineering will at our option repair or replace the defective device free of charge and the device will be returned...

Страница 19: ...3310 E 2015 Crystal Engineering Corporation 708 Fiero Lane Suite 9 San Luis Obispo California 93401 8701...

Отзывы: