Maintenance
— 48
SPECTROLAB — 23.07.2020
3. Remove the old filter mats.
4. Install the new filter mats.
5. Reinstall the covers.
Insert the covers from above and press against the top left and right
corners of the cover until it clicks into place.
Now, the filter mats are replaced.
7.8
Replacing the ICALization sample
When the ICALizing sample is almost used up, a replacement must be provided.
The composition of the ICALizing sample is different from that of the original
samples. The intensities of the new samples must be matched to those of the old
samples.
Please contact your Service Center.
Information
Do not discard the old ICALization sample before the intensities have been
matched. Matching is not possible without using the old samples.
7.9
Replacing the standardization sample
When the standardization sample is almost used up, a replacement must be
provided.
The composition of the new standardization samples is different from that of the
original samples. The intensities of the new samples must be matched to those of
the old samples.
Information
Do not discard the old standardization samples before the intensities have been
matched. Matching is not possible without using the old samples.