Serial Communications
|
5-9
Figure 5-7b.
Variable Table.
Name
ID
Hex
Description
Type
Typical Value
Moisture
91
5B
Moisture concentration reading
Float
20.0 ppm
HeldConc
92
5C
Held moisture concentration reading
Float
20.0 ppm
Frequency
93
5D
Sensor frequency reading
Float
1200 Hz
DeltaFrequency
94
5E
W et minus dry reading
Float
24.7 Hz
Flow3
97
61
Held and displayed flow
Float
50.0 sccm
ValveState
99
63
0-sample, 2-reference, 3-zero
Integer
0(sample)
Hour
100
64
Analyzer current time read
Integer
13
Minute
101
65
Analyzer current time read
Integer
36
Second
102
66
Analyzer current time read
Integer
10
Month
103
67
Analyzer current time read
Integer
12
Day
104
68
Analyzer current time read
Integer
5
Year
105
69
Analyzer current time read
Integer
2003
Date
106
6A
Analyzer current date
String
02/24/03
Time
107
6B
Analyzer current time
String
08:45
SystemState
108
6C
Alarm status word
Integer
1024
DataState
110
6E
Data ready variable
Integer
0
ModelName
119
77
Analyzer model name
String
3050-AM
ReferencePer
120
79
Current reference period time
Integer
15
SamplePer
121
79
Current sample period time
Integer
21
W etFreq
122
7A
W et frequency reading
Float
1254.7 Hz
DryFreq
123
7B
Dry frequency reading
Float
1230 Hz
W eekDay
124
7C
Current day of the week
Integer
1
MoistureSpan
125
7D
Concentration coefficient
Float
1.0
ZeroCycleCount
128
80
Zero cycles countdown
Integer
8
FrequencyOffset
131
83
Zero correction coefficient
Float
0.0
Font
156
9C
Display font flag
Integer
0
EquilZero
157
9D
Zero equilibration cycles set
Integer
3
OldFreqOffset
166
A6
Offset before last correction
Float
0.0
MoistureUnits
168
A8
Moisture units of measure set
Integer
0
Содержание 3050-AM
Страница 12: ...xii 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 22: ...2 4 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 3 Figures 3 1c Model3050AMdimensionstopand sides...
Страница 42: ...3 20 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 64: ...5 10 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 66: ...6 2 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 84: ...7 18 3050 AM Moisture Analyzer This page intentionally left blank...
Страница 96: ......