Installation & Start-Up
|
3-7
8
Power connector
Connect hot (1) and neutral (2) to power
connector and ground to chassis.
9
Fuse Receptacle
2A 250V, time lag, T-type.
10
Line Voltage Selector
Select 120 or 240 volt line voltage.
11
RS-485 Terminator
RS-485 termination switch is on if the
analyzer is the only analyzer on line or
the last in a chain.
12
RS-485 Serial Port
User serial communications interface.
Serial wiring supplied by user.
13
Analog Output
4 to 20 mA output to user recording
equipment. Wiring supplied by user.
14
Alarm Contacts
Connections for: Data Valid, Concentra-
tion Alarm and System Alarm. Wiring
supplied by user.
Figure 3-3.
Model 3050 AM interface
board connections.